TRUE METRIX Control Solution Care
• Write date opened on control solution bottle label.
Discard if either 3 months after first opening or after
date printed next to EXP on label has passed, which-
ever comes first.
• After each use, wipe bottle tip clean and recap tightly.
• Store at temperatures between 36°F-86°F.
DO NOT FREEZE.
TRUE METRIX Self Monitoring Test Strip Care
• Store test strips in original vial only. Do not transfer
test strips to new vial or store test strips outside of vial.
• Write date first opened on test strip vial label. Discard
unused test strips from vial if either 4 months after first
opening or after date printed next to EXP on label has
passed, whichever comes first. Use of test strips past
expiration dates may give incorrect results.
• Close vial immediately after removing test strip.
Store in a dry place at temperatures between
40°F-86°F. DO NOT FREEZE.
• Do not reuse test strip.
• Do not bend, cut or alter test strips in any way.
TMX_OB_GutsENG_RE4NPD03r6.indd 43
7/15/15 3:37 PM