Unpacking 

1.  To remove the Lid, hold the device firmly and slide the Lid as indicated on the 

top. 

 

2.  Pull  out  the  blue  Wedges  to  release  the  Inner  Chamber  (Cathode  Buffer 

Chamber Frame). 

3.  Take Out the Inner Chamber. 
 
 
 

Assembly and Operation 

Caution: Electrophoresis operates  at a high voltage and Gel Cassetes are fragile by 
nature. Some protective measure like wearing gloves should be taken when handling 
FastGene

®

 Protein Electrophoresis Chamber. 

 

 
1.  Ensure that the power supply is turned off and the Lid is not connected. 

2.  Ensure  the  Gaskets  are  set  on  the  Inner  Frame  (Cathode-Buffer  Chamber 

Frame).  Otherwise  place  the  Gaskets  on  the  grooves  of  the  Inner  Frame  and 
press the Gaskets so that they are fixed firmly into the grooves. 

3.  Pour the buffer (approximately 200 ml) in the Outer Anode Buffer Chamber up 

to around 4 cm from the bottom.  

4. 

Take two IDMax™ Precast Gel Cassettes and remove the combs gently

. Take the 

Inner Frame and position the Gel Cassettes so their bottom is covering all the 

“white”  part  of  the  gasket

  and  the  loading  opening  facing  toward  the  inner 

frame. 
If  using  only  one  gel  at  a  time,  use  one 

Dummy  Gel  Cassette

  instead  of  the 

second IDMax™ Precast Gel C

assette. Place the Dummy Cassette upside down, 

and  the  loading  opening  outside  of  the  central  reservoir,  so  it  will  not  allow 
contact between the 2 buffer reservoirs. 

 

Содержание FastGene FG-02

Страница 1: ...For Research Use Only For safety this manual must be read carefully and the device be handled properly FastGene Protein Electrophoresis Chamber Instruction Manual...

Страница 2: ...rent Limit 0 100 mA Ambient Temperature Limit 0 50 C Protective gloves and safety glasses must be worn when operating in a laboratory environment This device is intended for in vitro use only and must...

Страница 3: ...Gel Cassettes 6 o 1x Instruction Manual Specifications o Lid Acrylic Acrylic o Anode Buffer Chamber Frame Acrylic o Cathode Buffer Chamber Frame Acrylic o Gasket Silicone o Wedge Acrylic o Overall si...

Страница 4: ...athode Buffer Chamber Frame Otherwise place the Gaskets on the grooves of the Inner Frame and press the Gaskets so that they are fixed firmly into the grooves 3 Pour the buffer approximately 200 ml in...

Страница 5: ...ng of the level of the inside buffer If leakage is observed disassemble the inner Chamber and repeat the above steps Note Avoid a short circuit do not overfill the Inner Chamber 7 Fill the Outer Chamb...

Страница 6: ...ches approximately 5 mm around the center of Gasket from the bottom edge of the Gel Plate 14 Disconnect the power cables of the Lid for the power supply Holding the device firmly side the Lid as indic...

Страница 7: ...ne Protein Electrophoresis Chamber FG 02 IDSolTM Running Buffer 10x Running Buffer Tris Glycine SDS 500 ml ID1501 10x Running Buffer Tris Glycine 500 ml ID1511 10x Running Buffer Tris Tricine SDS 500...

Отзывы: