Remember
•
It is of vital importance to adhere to the specified operating conditions:
Ambient temperature: +16 to 30°C
Humidity: 20 to 60% RH (non-condensing)
• Mobile phones and cordless phones might interfere with the instrument
and should therefore be kept away from the instrument. Interference
could make it impossible to perform a measurement.
• When within the specified operating conditions it is recommended to
keep NIOX MINO
®
with installed Sensor connected to a power outlet
whenever possible.
• It is recommended to allow a new Sensor to be connected for two hours
before the first measurement is done.
• It is recommended not to use NIOX MINO in the proximity of areas
where volatile substances such as organic fluids or disinfectants are
being used. Special attention should be paid to aerosols and disinfection
baths (either open vessels or ultrasonic baths).
• Always use a NIOX MINO Bag (closed) for transportation and storage
of NIOX MINO when it is disconnected from the power outlet. (The
NIOX MINO Bag is sold separately.)
• Shelf-life - NIOX MINO instrument: Minimum 3 years at time of delivery,
or 3,000 measurements.
NIOX MINO Sensor: Maximum 12 months after opening package and
mounted in NIOX MINO, whichever comes first.
Note!
This manual is intended for medical care personnel and for use
when teaching patients how to use the NIOX MINO instrument.
To aid this, a demonstration mode is available in the instrument.
Please refer to the "Demonstration mode" section, page 7.