KSP01/00/12.04.2019
KSP01/00/12.04.2019
Guidance and manufacturer’s dec
l
aration – e
l
ectromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.The customer or the user of the
device should assure that it is used in such an environment.
Immunity test
EN 55014-1 Test
l
eve
l
Comp
l
iance
l
eve
l
El
ectromagnetic environment – guidance
Immunity to
Conducted
Disturbances,
Induced by
Radio
Frequency
Fields
IEC 61000-4-6
150 kHz to 230
MHz
3Vrms
80% Am at
1kHz Sinusoidal
1% step width 2 s
dwell time
150 kHz to 230
MHz
3Vrms
80% Am at
1kHz Sinusoidal
1% step width 2 s
dwell time
Portable and mobile RF communications
equipment should be used no closer to any part of
the device including cables, than the
recommended separation distance calculated from
the
equation applicable to the frequency the
transmitter.
Recommended separation distance:
80 MHz to 800 MHz
800 MHz to 2.7 GHz
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.Interference may occur in the
vicinity of equipment marked with the following
symbol:
Note 1:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures,objects and people.
39
Recommended separation distances between portab
l
e and mobi
l
e RF communications equipment and the device
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz
(out ISM and
amateur
radio bands)
150 kHz to 80 MHz
(out ISM and
amateur
radio bands)
80 MHz to 800 MHz
800 MHz to 2.7 GHz
0.01
0.12
0.06
0.12
0.23
0.1
0.38
0.19
0.38
0.73
1
1.2
0.6
1.2
2.3
10
3.8
1.9
3.8
7.3
100
12
6
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W)according to the transmitter manufacturer.
Note 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2:
The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to
6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The
amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz
to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz,21,0 MHz to 21,4
MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
Note 3:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures,objects and people
.
Guidance and manufacturer’s dec
l
aration – e
l
ectromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.The customer or the user of the device
should assure that it is used in such an environment.
Immunity Test
EN 55014-1
Test Leve
l
Comp
l
iance
l
eve
l
El
ectromagnetic environment –
guidance
Discontinuous
Disturbance / Click
Test
150 kHz, 500 kHz, 1.4
MHz, 30 MHz
150 kHz, 500 kHz, 1.4
MHz, 30 MHz
Device is supplied by LISN equipment on
L, N, PE couplings in the shielded room.
When device is in the normal priod of
operation click value that was given to
main supply is measured by EMI Test
Receiver.
Conducted
Emission
150 kHz – 30 MHz,
Quasi-Peak, Average
Detector.
150 kHz – 30 MHz,
Power Disturbance
Test
30 MHz – 300 MHz
30 MHz – 300 MHz
Device, has been placed on a wooden
isolator which was 80 cm height from
referance ground plane in the shielded
room. It is tested to find the voltage value
to produce the maximum emission on the
freqquncies 30 MHz and 300 MHz supply
voltage.
40
Содержание HNK-SUTPUMP-01
Страница 23: ......