background image

Digital Microscopy Center, University of Washington

 

July 29, 2016   

 

 

 

 

 

 

 

 

 

5

 

 

A. Performing Brightfield transmitted light microscopy 
1.

 

Focus on specimen 

1.1.

 

Place slide on the stage and hold in place with the stage clip

 

1.1.1.

 

Begin with low magnification for greater ease in surveying your specimen to 

identify regions of interest 
 

2.

 

Adjust the focusing eyepiece 

2.1.

 

Find a small object with sharp boundaries

 

2.2.

 

Use the fine focus using observing with one eye

 

2.3.

 

rotate the upper portion of the other eye piece until the object is in similar focus with both 
eyes

 

 

3.

 

Set Koehler Illumination for objectives 10X and above (Figure 1, Figure 2) 

3.1.

 

Note: this should be performed every time the objective is changed 

3.2.

 

Focuses the condenser to the same plane as the objective, set the incoming light column and 
condenser NA 

3.2.1.

 

This is essential for best contrast, resolution and even illumination 

3.3.

 

Open the condenser diaphragm for maximum brightness 

3.4.

 

Make certain the condenser top lens is flipped up 

3.5.

 

Close the field stop diaphragm to its minimum diameter 

3.6.

 

Use the condensor focus knob to bring the leaves of the field stop sharply into focus while 
observing through the eyepieces 

3.7.

 

Roughly center the spot of light with the centering screws 

3.8.

 

Open the field stop until it just reaches the edge of the field of view 

3.9.

 

Adjust the centering until it is concentric with the field of view 

3.10.

 

Close the condenser diaphragm to match the NA of the objective 

3.10.1.

 

If the objective NA is higher than .85, set condenser to maximum 

3.11.

 

Double check the eyepieces – if the image is blurry, close the field stop, re-focus condenser 
then re-open the field stop 
 

4.

 

Set Koehler Illumination for objectives below 10X 

4.1.

 

Follow the steps above, except that after step 3.7, flip the condenser top lens out of position 

 

5.

 

Setting contrast filters 

5.1.

 

A color filter inserted into the light path, as paddles at the lamp housing or placed on the 
condenser, can increase contrast. 

5.1.1.

 

Viewing blue or brown staining often benefits from using a green or blue filter 

 
 
 
 
 
 
 
 

Содержание Optiphot

Страница 1: ...the trinocular beamsplitter to direct the image to the eyepieces 2 Brightfield 2 1 Turn on the microscope lamp 2 1 1 black push button at left front base of microscope 2 2 Select your objective lens 2 3 Set the trinocular beamsplitter to direct the image to the eyepieces 2 4 Select DIA on filter slider 2 5 Adjust lamp intensity with slider across base of microscope 2 6 Focus on the specimen 2 7 Se...

Страница 2: ...7 If the bulb fails to ignite it may be too hot from prior use the bulb may be faulty or may need to be reseated Contact facility staff 2 Transmitted Illumination with halogen lamp Figure 1 2 1 Power switch turned on by the black switch left front corner of the microscope base 2 1 1 This lamp is used for brightfield phase contrast and darkfield microscopy 2 2 Lamp intensity is controlled by 3 mean...

Страница 3: ... Filter Light Reduction Transmission ND2 1 2 50 ND4 1 4 25 ND16 1 16 6 25 ND2 ND4 1 8 12 5 ND2 ND16 1 32 3 125 ND4 ND16 1 64 1 562 ND2 ND4 ND16 1 128 0 781 Condenser focus Substage field diaphragm Brightfield lamp power Brightfield lamp voltage Figure 1 Sub stage Controls Fine Focus Coarse Focus Transmitted Light ND Filters Photo switch ...

Страница 4: ...and weak fluorescence but large working distance 4 1 2 low magnification providse large field of view but may not magnify sufficiently for full resolution at the camera Locking screw for stage rotation Condenser diaphragm NA adjustment Centering screws for condenser alignment Lever to flip condenser top lens in or out of optical path Stage Y axis motion Stage X axis motion Figure 2 Sub stage Conde...

Страница 5: ...ximum brightness 3 4 Make certain the condenser top lens is flipped up 3 5 Close the field stop diaphragm to its minimum diameter 3 6 Use the condensor focus knob to bring the leaves of the field stop sharply into focus while observing through the eyepieces 3 7 Roughly center the spot of light with the centering screws 3 8 Open the field stop until it just reaches the edge of the field of view 3 9...

Страница 6: ...d DAPI cubes are swapped see DMC staff for instruction on changing the 4th filter 6 3 Xenon lamp intensity is controlled by pushing ND filters in or out of the light path 6 3 1 located behind the trinocular head Figure 3 6 4 Xenon excitation is controlled by a slider located between the ND filter and the xenon lamphousing Figure 3 6 5 The excitation light is also blocked when the filter slider is ...

Страница 7: ...signated as 160 do not require a coverslip Lenses designated as 160 17 require a coverslip of 170 µm thickness Filter Label Cube Label Color Part Nikon FITC HQ FITC green 96170 C01174 TxR HQ TXRD red 96172 C00178 UV UV 2E C blue blue Cy5 Cy5 far red 31023 C81736 Lens µm pixel pixel µ m 1 mm reticle field 1 04 Plan 160 8 40 0 12 2 05 Plan 160 4 24 0 24 4 1 EPlan 160 2 12 0 47 10 40 EPlan 160 0 87 1...

Отзывы: