Maintenance Service on Nikon CoolScan Slide Scanner 

 

Summary: Major disassembly of scanner mechanism and controller circuit 
board, lubricate and manually actuate slide rails and stepper motors.  
Time to perform: about 2 hours 

 
The X-axis scanning runs on three slide rails and a stepper motor. The Z-axis Autofocus runs on 2 
slide posts, 2 forked sliders and a stepper motor. Both of these mechanisms require lube until they 
move easily and smoothly. Circuit board removal is required as most of the mechanism to be lubed 
is on the left side. Only one X-axis slide rail is accessible from the right side, making it an 
incomplete job. Clean sensor lens and two front-sided mirrors, as needed. 
 
Statement of Warranty, Limits of Liability agreement, and other WARNINGS is included in 
Section IV, near the end of this document. 
 

I. Qualify that the unit is operational, incoming inspection 

 
0) check to see that no data cable is attached and that jacks are not damaged or contact pins are bent 
or corroded. 
 
1) make sure that unit turns on with just AC plugged in. See that green light turns on, blinks and 
stays steady. Position unit in vertical position with power switch on top. 
 
2) during the blinking phase, the unit is booting up and goes through mechanism qualification 
steps, see if you can hear any buzzing where stepper motors are moving the scanner mechanism. 
Testing of Autofocus is first and X-axis scan is second. 
 
3) look inside the slide module throat and see if you can see any narrow strip of light from the LED 
light module located in the upper case near where the inner edge of the slide mount would be.  
 

Symptoms: 

 
4) if any of the mechanisms doesn't move during the qualification step, the green light changes to a 
rapid blink (about every half sec) and then steady green at the end of the boot sequence. 
 
This negative failing rapid blink diagnostic will show up by the NikonScan app not recognizing 
that a correctly booted scanner is attached...the infamous "Nikon Scan was unable to find any 
active devices" dialog box message.  
 
If yes to all four, then the unit just needs to be lubricated with Teflon and lens and mirrors cleaned 
with alcohol and or aqueous lens cleaner. 
 
5) if there is slow blinking (about once per second) during boot for about 15 sec and then a steady 
green, then the slide scanner has booted normally and this maintenance and cleaning procedure is 
not required at this time. 

Содержание LS-2000

Страница 1: ...s steady Position unit in vertical position with power switch on top 2 during the blinking phase the unit is booting up and goes through mechanism qualification steps see if you can hear any buzzing where stepper motors are moving the scanner mechanism Testing of Autofocus is first and X axis scan is second 3 look inside the slide module throat and see if you can see any narrow strip of light from...

Страница 2: ...d so precisely that screws seem to stick on the tip without being magnetic May need a medical type hemostat forcep or tweezer to pick out dropped and re position tiny screws Available at an unusually complete Ace hardware store such as Seven Corners Hardware St Paul Minnesota 651 224 4859 7corners com with an upstairs nationally rated catalog mail order operation at 800 328 0457 2 DuPont Teflon Mu...

Страница 3: ...flashlight 9 6v Ni Cd with ratcheting right angle head or equivalent Perhaps a Petzl Tikka forehead strapped LED flashlight will work too 4 Plastic pink antistatic foam pad 1 ft x 18 x 1 2 thick preferably grounded with 2 Meg ohm resistor 5 Cheseborough Ponds cotton Q tips or TexWipe foam swabs and aqueous lens cleaner isopropyl alcohol or Everclear liquor 190 proof ethanol Kimberly Clark Chemwipe...

Страница 4: ...connectors put in vial 2 The circuit board should be loose but still attached E remove 5 cables on bottom side of circuit board and 2 cables on top side of circuit board May need to use fingernail thin jewelers screwdriver or forcep to separate the cable jack and connector Refrain from pulling on the wires with fingers or forcep just get it to barely separate then insert and twist a flat bladed sc...

Страница 5: ...eft side especially for any bloated or leaking round cylindrical brown capacitors 4 larger ones 3 black striped smaller ones There is a soldered in ceramic 3 15A internal fuse F1 Disassemble unsolder and replace if needed J remove front plastic bezel there are top and bottom plastic snaps K lubricate the three X axis rails with Teflon with several drops of lube Use a flat bladed screwdriver and su...

Страница 6: ...the mechanism back and forth from end to end When satisfactory return the carriage to a central position If a medical syringe is available it may be easier but the Teflon may settle out and the vapor pressure of carrier vehicle may cause it to squirt out on unintended places N On the bottom of the mechanism there is a spring loaded brown bronze clip spring covering the Autofocus Z axis stepper mot...

Страница 7: ... scanning assy is moved back and forth Time to test the bare mechanism under live power be careful to not touch power supply parts which may have dangerous high voltage Plug in power cord and turn on power switch S watch the shaft of Z axis Autofocus stepper motor during the beginning part of the boot process it must rotate back and forth The carriage can be manually moved out of mid range in both...

Страница 8: ... in elongated sliding slots Fasten in the mid position IV Warranty and Limits of Liability This procedure is provided on a best efforts basis and there is no warranty whatsoever herein and is provided AS IS It requires a high level of technical competence and mechanical dexterity such as an A Certification Computer Technician plus 3 years experience or equivalent and demonstrated ability to handle...

Страница 9: ... chip Firmware 30 soldered in main chips Nikon EI 104 vE17 main controller Hitachi HD6413003F16 H8 3003 microcontroller Analog Devices AD9220AR 12 bit A D converter Analog Devices AD7528JR 8 bit D A converter LSI Symbios 53CF92A SCSI controller WinBond W24257AJ 15ns memory x8 JRC 7809A voltage regulator high quality capacitors are metal encased CCD Signal Processing board on sensor assy P N F17045...

Отзывы: