background image

Nikon Imaging Centre @ King’s College London

 

Adjusting K

ӧ

hler illumination (Transmitted light) 

For  transmitted light  imaging it is essential to optimise  the condenser alignment  to ensure  that the full numerical 
aperture of the microscope  system can be realised (to provide an even illumination of the sample). It is therefore 
important to ensure that the light passes through the condenser optimally to allow for high quality imaging.  

N.B. K

ӧ

hler Illumination adjustment should only be conducted through the eyepiece. 

 

1.

 

Through the eyepiece, focus on your. Fully close the field iris by turning it anticlockwise. (Fig. 14)

 

2.

 

Using  the  adjustment  knob,  move  the  condenser  whilst  looking  through  the  eyepiece  until  you  see  the 
octagonal field iris. Turning the knob towards you to bring the condenser lens close to the sample. Adjust the 
condenser focus so that the outline of the field iris is sharp and in focus. (Fig. 15) 

 

3.

 

If required, centre the field iris in the eyepiece using the adjustment screws on the condenser unit.

 

4.

 

Once centred, open the field iris again so that the field iris edges are just outside of the field of view.  

 

 

 

Nikon 

Imaging Centre 

King’s College London 

Field stop iris 
diaphragm 

Alignment screws (use 
allen key if centering 
required) 

Condenser 
focus  

Fig. 13

Condenser 
lens  

Содержание Live Cell 2

Страница 1: ...Live Cell 2 guide NIKON IMAGING CENTRE KING S COLLEGE LONDON King s College London Nikon Imaging Centre ...

Страница 2: ...r joystick The speed of movement is indicated on the joystick display panel denoting slow movement 3 Objective Lenses on the live cell system are 4x 10x 20x and 40x dry objectives Check the correction collar is set to 0 17 on the 20x and 40x objective lenses for imaging through standard 1 5 coverglass the collar can be adjusted for thicker samples 60x and 100x objectives can be added to the system...

Страница 3: ...don Nikon Imaging Centre King s College London Objective selector Light path Fig 5 Intensity dial Transmitted light on off Condenser element selector Fig 6 Fig 7 Epi fluorescence shutter Epi fluorescence filter turret Fig 8 PFS on off z speed ...

Страница 4: ...s accompanied by an audible alarm which can be silenced by pressing the alarm indicator on the touch screen 2 Selecting temperature or CO2 opens a second window where the set point can be adjusted Fig 13 The values can be changed using the buttons Confirm the settings by pressing set 3 For CO2 the valve on the tank needs to be adjusted to ensure that it is entering the system correctly open the to...

Страница 5: ...4 Nikon Imaging Centre King s College London Fig 11 Fig 12 Nikon Imaging Centre King s College London Fig 13 Press to set CO2 concentration ...

Страница 6: ...cus on your Fully close the field iris by turning it anticlockwise Fig 14 2 Using the adjustment knob move the condenser whilst looking through the eyepiece until you see the octagonal field iris Turning the knob towards you to bring the condenser lens close to the sample Adjust the condenser focus so that the outline of the field iris is sharp and in focus Fig 15 3 If required centre the field ir...

Страница 7: ...f the configurations from the eyes group of configurations These configurations 1 Set the light path to EYE 2 Set the filter in the turret to that relevant to the dye being used 3 Make sure that the fluorescence lamp is turned on and use the microscope base for control of the fluorescence shutter 4 Transmitted light imaging can be achieved using the relevant OC and using the lamp on off but on the...

Страница 8: ...7 Acquiring a colour image DS U3 camera 1 Select the colour camera BF optical configuration from the OC Panel Fig 16 2 Focus on the sample and adjust the detector settings so that you can see the surface 3 Set the transmitted light to 100 4 Move to a blank region of the sample Fig 19 and select Auto White in the DS Fi2 setting window Fig 18 This will ensure that your images have the correct white ...

Страница 9: ...y if these means setting a very long exposure time 4 The PE pad can be used to adjust the intensity of the LED lamp This defaults to 100 but can be reduced if the sample is prone to photobleaching or phototoxicity 5 The camera used for fluorescence imaging has two resolution options which vary the number of pixels in the image This can be changed with the channel settings however if a multicolour ...

Страница 10: ...which will require the user to stop the acquisition after the desired period 3 Other ND acquisition modes can be added to the timelapse e g z stacks multipoints which when checked and set up in the ND acquisition window are added to the image sequence NOTE multidimensional parameters will affect the achievable frame interval of the timelapse Clicking timing will allow you to see if the interval is...

Страница 11: ... menu bar 5 Select the objective for capturing the image set the number of fields in XY and whether the tiling should scan around the current position of the stage or with the stage position being the top left Set the overlap to 15 and select blending making sure image registration is checked Fig 27 6 Large image files can either be saved as the whole large image the single frames as TIFFs or both...

Страница 12: ... any previously saved z positions 2 Three options are available for setting Z stack limits Top and Bottom mark the limits of the stack Symmetric mark the centre of the stack and Asymmetric set the number of slices above and below the centre 3 For the top bottom option whilst live imaging adjust the focus to find the top of your sample use the Z value as a guide and click Top Focus back through the...

Страница 13: ...drive E and at the end of the session transfer your files to the network storage drive N LiveCellData Data is kept for a maximum of 1 week in the documents folders and 1 month on the network drive 3 Instructions on how to connect the network drive to your King s computer can be found when you login to your PPMS account go to documents and select the N drive access document Reusing settings 1 When ...

Страница 14: ...xt hour leave all the components on except for the CO2 if you have used it and just exit the software 2 Remember to transfer your data from the datadrive to livecelldata 3 Remove your sample from the microscope stage 4 Lower the objective fully using the manual focus wheel and move the stage to the central position 5 If no user is due on the system after you shut down all components in reverse ord...

Отзывы: