6
Checking and Adjustment
132
Total Station DTM-322 Instruction Manual
Checking and Adjusting the Plate Level
The axis of the plate level vial must be at right angles to the vertical axis of the
instrument.
To check and adjust the plate level:
1.
Set up the instrument on the tripod.
2.
Follow the leveling procedures described in
3.
Rotate the alidade 180°.
4.
Check whether the bubble is in the center of the vial.
5.
If the bubble is not in the center of the vial, adjust the plate level:
a.
Using the adjusting pin supplied, rotate
the plate level’s adjustment screw until
the bubble has moved half of the
distance back to the center.
b.
Using leveling screw A, move the
bubble into the center of the vial.
c.
Repeat from
.
Checking and Adjusting the Circular Level
Once you have checked and adjusted the plate
level, check the circular level.
If the bubble is not in the center of the level, use
the adjusting pin to rotate the three adjustment
screws until the bubble is centered.
C
B
A
2
Содержание HQA46480
Страница 1: ...www trimble com Total Station DTM 322 Instruction Manual Version A 1 0 0 Part Number C232E April 2009 ...
Страница 8: ...viii Total Station DTM 322 Instruction Manual ...
Страница 14: ...Contents xiv Total Station DTM 322 Instruction Manual ...
Страница 60: ...3 Getting Started 46 Total Station DTM 322 Instruction Manual ...
Страница 157: ...C H A P T E R 8 Total Station DTM 322 Instruction Manual 143 System Diagrams 8 In this chapter System Components ...
Страница 160: ...8 System Diagrams 146 Total Station DTM 322 Instruction Manual ...
Страница 176: ...9 Communications 162 Total Station DTM 322 Instruction Manual ...