Chapter 2
Troubleshooting
33
T
roub
le
shooti
ng
2
Electrical Requirements
2.1
General
■
Power
Problem
Cause
Measure
There is no power even
though the power switch is
on.
The power cord is not connected, or is
connected improperly.
Connect the cable properly.
(
→
See Chapter 1, “3 Assembly Method, 20 Connect the
motorized device cable” in the “Assembly/Maintenance”.)
A contact arm is not used.
Use a contact arm.
(
→
See Chapter 1, “3 Assembly Method, 5 Attach the
standard arm/contact arm” in the
“Assembly/Maintenance”.)
The motorized unit connected to the contact arm
is attached incorrectly.
Attach the motorized unit correctly.
(
→
See Chapter 1, “3 Assembly Method” in the
“Assembly/Maintenance”.)
A motorized accessory
does not operate.
The motorized unit connected to contact arm is
not in correct combination.
Attach the motorized unit in the correct combination.
(
→
See Chapter 1, “3 Assembly Method - Introduction” in
the “Assembly/Maintenance”.)
Contact arm LED is
blinking.
The motorized unit connected to contact arm is
not in correct combination.
Attach the motorized unit in the correct combination.
(
→
See Chapter 1, “3 Assembly Method - Introduction” in
the “Assembly/Maintenance”.)
■
Illumination
There is no power supplied.
Plug in the power cord.
(
→
See Chapter 1, “3 Assembly Method, 21 Connect the
power cord” in the “Assembly/Maintenance”.)
The lamp has burned out.
Replace the lamp with the specified type.
(
→
See Chapter 3, “1 Replacing the Lamp” in the
“Assembly/Maintenance”.)
Lamp does not light.
The lamp is not attached.
Attach a designated lamp.
(
→
See Chapter 1, “3 Assembly Method, 3 Attach the lamp”
in the “Assembly/Maintenance”.)
■
Controls on the main body/controller
Problem
Cause
Measure
Brightness cannot be
adjusted with the
dia-illumination brightness
control knob.
Operation is disabled.
Change the setting.
(
→
See Chapter 3 “1.1 Adjustment by Lamp Voltage”
-
Turning lamp ON, brightness adjustment control” in
the “Operation”.)
A contact arm is not used.
Use a contact arm.
(
→
See Chapter 1, “3 Assembly Method
- 5 Attach the
standard arm/contact arm” in the
“Assembly/Maintenance”.)
The simple remote control
pad does not work to move
the motorized unit.
Cables for the contact arm, control box, and the
simple remote control pad are not connected
properly.
Connect the cable properly.
(
→
See Chapter 1, “3 Assembly Method
-
20 Connect the
motorized device cable” in the “Assembly/Maintenance”.)