Niked Medical Mark 5 Nuvo M5C5 Скачать руководство пользователя страница 6

2010-2224F Rev C

6

May 2011

MARK 5 Nuvo [M5C5]

Serial No. ____________

Date  rst used: 

___________________________

_________________________________________

Maintained by: ___________________________
_________________________________________

Your distributor: _________________________

Address : ________________________________
_________________________________________

Telephone : ______________________________

PREVENTIVE MAINTENANCE:

a

. Wash cabinet dust  lter weekly.

b. Inspect inlet air  lter at each patient visit. 

    Replace  lter annually, or more often depending on 

environment.

c. Check oxygen concentration every 5,000 hours or

one year.

The manufacturer’s instructions for the

preventive

maintenance

of the devices are de ned in the

maintenance manual. Check with your service provider
for any updates to the recommended schedules.

The work must be carried out by suitably trained

technicians certi ed by the manufacturer.

Use original spare parts only (see Pg. 7).

Upon request, the supplier can provide circuit diagrams,
spare parts lists, technical details or any other information

of  use to quali ed technical personnel for parts of  the 

device which are designated as being the manufacturer’s
responsibility or by the manufacturer as repairable.

Medical Device Regulations require users and
service providers to report to the manufacturer any
incident that could, if repeated, result in injury to

Electrical power supply:

115 V Units

230 V Units

Frequency:

60Hz

50/60Hz

Average Power:

410 W

420 W

Protection Class:

ClassII

Class II

Mains Protection:

10A

5A

Filters:

At the rear of the device: a cabinet dust  lter.

At the compressor  input:  an inlet air  lter,  behind cabinet 

air  lter.   

Before  the  oxygen  outlet:  a   nal  product   lter  <  0.3  µm. 

(technician only).

Air circulation:

A tubeaxial fan cools the compressor compartment
.

Environmental limit conditions:

The performances of the device (especially the oxygen
concentration) are quoted at 21

o

C (70

o

F) and one atmosphere.

They may change with temperature and altitude. For further
information, please consult the maintenance manual.
• The device must be stored, transported and used in the
vertical position only.
• Ambient temperature of between 5

o

C and 40

o

C (40

o

F and

100

o

F) (operation).

• Storage temperature range -20

o

C to 60

o

C. (0

o

F to140

o

F).

• Relative humidity of between 15 % and 95 % operation and
less than 95% storage, both non-condensing.
• Altitude(21

o

C): Up to 1500 m (5000 ft) without degrada-

tion;
Consult your equipment provider for further information
regarding 1500 m to 4000 m (5000 to 13000 ft).
• IPX1: Complies with EN60601-1 standard; spilling of a
glass of water.

IV. 7. Standards

EN 60601-1[UL60601-1:2003],CAN/CSA-C22.2 No.601.1-
M90 w/A1&A2: Electrical Safety- Medical Devices.
EN60601-1-2:2000 Electromagnetic Compatibility

I

V. 8. Method for disposing of waste

All waste from the

MARK 5 Nuvo [M5C5]

(patient circuit,

 lter, etc.) must be disposed of using the methods appropriate 

to the civil authority of the location where disposed.

IV. 9. Method for disposing of the device

In order to preserve the environment, the concentrator must
only be disposed of using the appropriate methods. All ma-
terials of construction are recyclable.

Cabinet
Dust
Filter

Air
Filter

Содержание Mark 5 Nuvo M5C5

Страница 1: ...t function 5 IV 6 Technical characteristics 5 IV 7 Standards 6 IV 8 Method for disposing of waste 6 IV 9 Method for disposing of device 6 CONTENTS ON power switched on Off power switched off Type B de...

Страница 2: ...acturer assembler installer or distributor are not considered to be responsible themselves for the consequences on the safety reliability and characteristics of a device unless Theassembly tting exten...

Страница 3: ...he correct voltage and frequency as de ned on the manufacturer s technical label 8 f Press the power switch I 0 to the ON position I The red indicator will light and the audible alarm will sound for a...

Страница 4: ...o M5C5 istobecleaned with a soft dry cloth or if necessary a damp sponge then thoroughly dried with wipes and an alcohol based solution Acetone solvents or any other in ammable products must not be us...

Страница 5: ...tlet and is calibrated to 2 7 bar 40 psig IV 5 3 Maintenance of the system alarms No special maintenance is required The equipment supplier checks that the unit is still operating correctlywhentherout...

Страница 6: ...Protection Class ClassII Class II Mains Protection 10A 5A Filters At the rear of the device a cabinet dust lter At the compressor input an inlet air lter behind cabinet air lter Before the oxygen out...

Страница 7: ...safety de vice has been activated Fan is not working Dirty Filters Tube disconnected or humidi er not tight Cannula tubing is kinked Observations The I O ON OFF button is in the ON position but the de...

Страница 8: ...2224F Rev C 8 May 2011 Nidek Medical Products Inc 3949 Valley East Industrial Drive Birmingham Alabama 35217 U S A Tel 205 856 7200 Fax 205 856 0533 MMainain Main Maintenance This i Maintenance Items...

Отзывы: