INSTALLATION PROCEDURE: Unpacking
4
Remove the plastic bag from the main body.
5
Remove the brackets (n = 4) that hold the
adjuster feet (n = 4).
1) Loosen the upper nut of the adjuster foot
from the bracket.
2) Remove the bolts (n = 2) and washers (n
= 2) to remove the bracket.
6
Put the main body on a place to install.
WARNING
• To transport the system, lift it with four or more persons, and put it on a cart.
You may hurt your back or may fall down if you do it alone. The main body weights
approx. 120 kg.
CAUTION
• Leave a space of 10 cm or more behind the system.
If the cooling fan is blocked, the temperature inside the system will rise and it may
cause malfunction.
• Install the system on the special table (option) or a table that is adequately strong
for the weight (approx. 120 kg).
Vibrations and noises may be produced, and normal grinding may not be performed.
7
Remove the tape that holds the grinding
chamber cover.
8
Open the grindng chamber cover, and
remove the cushion from the grinding
chamber.
9
Remove the cardboard from the space at
the lower right of the front.
Upper nut
Bracket
Bolt, Washer
Cushion
Tape
Cardboard