PARAMETER SETTINGS AND MAINTENANCE: Parameter Settings
Parameter option
Setting contents
91. STATION I/F
(Station interface)
CABLE / WLAN / IR / NO
Sets the communication method between the main body and station
•
CABLE: A connection cable is used for communication.
•
WLAN: Wireless LAN is used for communication.
•
IR: Infrared is used for communication.
•
NO: Communication does not occur between the main body and station.
* When the
parameter is set to “LAN”,
“IR” and “NO” cannot be selected.
* While the station is connected with the cable, wired communication is
established even when WLAN or IR is selected. (Note that the
parameter is set to “AD HOC” for the WLAN
setting.)
92. OUTPUT TO PC
(Data output to computer)
USB MEM. / WLAN / LAN / NO
•
USB MEM.: Data is output to the connected USB flash drive.
•
WLAN: Data is output to the computer by a wireless LAN.
•
LAN: Data is output to the computer by a LAN cable via station.
•
NO: Data output does not occur.
* When the
parameter is set to “IR” or “NO”,
“LAN” cannot be selected.
* When the
parameter is set to “WLAN” and
the
parameter is set to “INFRA.”, “LAN”
cannot be selected.
93. OUTPUT TO RT
(Data output to refractor)
YES / NO
Sets whether to output serial data to the refractor or Eye Care card.
* When the
parameter is set to “NO”, data
cannot be output.
94. FORMAT FOR RT
(Serial communication format)
ALL / SHORT
Sets the serial data output format for the refractor.
•
ALL: All data are transmitted.
•
SHORT: Only median values are transmitted.
95. RING IMAGE
(Ring image export)
YES / LOW CONF / NO
Sets whether to export the image data (JPEG) of a ring image.
•
YES: Image data is transmitted.
•
LOW CONF: When the confidence index is 7 or lower, image data is trans-
mitted.
•
NO: Image data is not transmitted.
* When the
parameter is set to “NO”,
data cannot be output.
96. RETRO IMAGE
(Retroillumination image
export)
YES / NO
Sets whether to export the image data (JPEG) of a retroillumination image.
* When the
parameter is set to “NO”,
data cannot be output.