DT-5TXR-FVTR/DT-5TXR-FVC (continued)

Instructions to set the analog output characteristics are

as follows:
1. Press the FUN button eleven times; the display will reflect:

Displayed is the minimum RPM (factory setting).

4. Using  the  numeric  buttons,  enter  the  minimum

number of RPM to be monitored.
For example: if using the 0 to 10VDC output and the

minimum RPM to be monitored is 1,000, then 1 0 0 0

would have to be entered; this will provide a 0VDC

output at 1,000 RPM.

5. Press the FUN button again; the display indicates:

The display reflects the factory setting.

6. Use  any  of  the  lower  buttons  to  toggle  between

the two settings for the update time for the analog

output.  “0”  means  that  the  update  time  will  be

approximately 10 msec, “1” indicates that the update

time will correspond with the display update time

already chosen (see “Setting Mode Parameters” #4).

7. Press the FUN button; display will reflect:

The display shows that the maximum RPM is set for

1,000 (factory setting).

2. Using  the  NUMERIC  buttons,  enter  the  maximum

number of RPM to be monitored.
For example: if using the 0 to 10VDC output and the

maximum RPM to be monitored is 1,500, then 1 5 0 0

would  have  to  entered;  this  will  provide  a  10VDC

output at 1,500 RPM.

3. Press the FUN button; the display will show:

The display shows the factory setting.

8. Use the lower buttons to program the desired offset

percentage  setting  for  the  analog  output.  This

number can be from –100% to 100%.

9. Press  the  SET  button  to  store  the  analog  output

characteristics and exit.

www.calcert.com

[email protected]

1.888.610.7664

0

5

10

15

20

25

30

Содержание Shimpo DT-5TS

Страница 1: ...Instruction Manual DT 5TXR DT 5TS DIGITAL PANEL MOUNT TACHOMETER www calcert com sales calcert com 1 888 610 7664 0 5 10 15 20 25 30...

Страница 2: ...s 3 Important Safety Precautions 3 Display and Keypad 4 Rear Panel and Terminals 5 Set Up 5 Power 5 Sensor Connection 5 Self Test 5 Panel Installation 6 Factory Settings 6 Parameter Settings 6 Functio...

Страница 3: ...mounting while saving time Accepts modules DT 5TXR only which provide different types of outputs i e analog voltage analog current 4 20 mA relay closures set points BCD output and others for recordin...

Страница 4: ...e t e m a r a p s t e S n o t t u B t e S 7 n o i t c e l e s e d o M n o t t u B e d o M 8 l a c e d r o f a e r A l a c e D t i n U 9 t n e m e r u s a e m f o s t i n u s y a l p s i D y a l p s i...

Страница 5: ...imal points etc 4 When satisfied that all LED s are operational press the SET button to exit SENSOR TYPE TERMINAL NUMBER FREQUENCY or RPMRANGE OPERATION TEMPERATURE FILTER M P R 0 0 0 5 0 C 0 6 B 1 E...

Страница 6: ..._ 01000 1 000 RPM 3 Display units 1000 1 000 4 Update time _ 1 0 _ 1 second 5 Display Hold after input signal ceases _ _006 0 6 seconds 6 Input filter _ 10 _ 10 kHz DEFAULT FACTORY SETTING PARAMETER...

Страница 7: ...d and programmed into the DT 5TXR DT 5TS In addition there are also function settings to further customize the DT 5TXR DT 5TS NOTE Because the flow rate measurement mode 04 is only used in the Japanes...

Страница 8: ...ding is feet per minute and the belt that is being measured is moving at a rate of 55 5 f min then 5 5 5 would have to be entered and the negative decimal point button pressed once If the desired read...

Страница 9: ...etting 7 Press the PAR button 8 Use any of the lower buttons to scroll through see table above and select an update time 9 Press the PAR button 10 Use the NUMERIC buttons to enter the time in seconds...

Страница 10: ...ement The meter will then use this number to set parameters 1 and 2 Set the teaching feature as follows 1 Determine the output RPM 2 Confirm that the meter is receiving a signal from the sensor the SI...

Страница 11: ...o s i s e r e t s y H 2 _ __ _ 0 0 0 3 Secures the 1st high and 1st low setting 0 not secured 1 secured _ 0 _ Notsecured d n o c e s 9 9 0 t i m i l t e s y n a f o n o i t a v i t c a t u p t u O 4 _...

Страница 12: ...Simply activate the MEM button by pushing it momentarily The lower left secondary display will reflect the maximum value of the main display and the lower right secondary display will reflect the min...

Страница 13: ...speed and the operator wants to observe the changes in this case higher update times are selected or if the speed changes rapidly within certain limits this may bother the observer a lower update tim...

Страница 14: ...DT 5TXR DT 5TS has a MODE for this application and similar ones where it can measure the ON time of a signal that starts when the process begins until it ends see signal waveform diagram top right NOT...

Страница 15: ...C and 4 20mA DOP CPTR Relay output HI GO LO 3 C type relays DOP TRTR NPN OC output H 1 H 2 GO L 1 L 2 and ZERO DOP RMTR Ratio of two input functions Lower Connector Options with Connectors C DOP FVC A...

Страница 16: ...VTR or DOP FVC module coupled with the DT 5TXR panel meter provides three popular voltage outputs and one current output for recording purposes and PLC applications The voltages are a 0 to 10 VDC b 1...

Страница 17: ...time will be approximately 10 msec 1 indicates that the update time will correspond with the display update time already chosen see Setting Mode Parameters 4 7 Press the FUN button display will refle...

Страница 18: ...DOP BCD Module DT 5TXR BCD The model DT 5TXR BCD combination satisfies PCs and PLCs that require a BCD parallel signal The DOP BCD module can be configured with a negative logic output 0 HI and 1 LO o...

Страница 19: ...x frequency 30 kHz 0 999999 with selectabledecimal point 0 008 1 digit DT 5TXR Display Range MODEL Mode r e t e M e t a R r e t e m o h c a T Elapsed Time Process Time 0 00 00 9 59 59 0 00 00 0 59 59...

Страница 20: ...ts INDEMNIFICATION LIMITATION OF DAMAGES Buyer agrees to indemnify and hold Shimpo Instruments harmless from and against all claims and damages imposed upon or incurred arising directly or indirectly...

Отзывы: