     4.5.6 Date/Time

The system time may be set under this menu. Touch “ZERO”       
to adjust the value. Press “MODE” to shift to the next position. 
Touch “LOG” to cancel. Press “MENU” to confirm and exit. 
(Figure 4-5h)

 

 

 

    Figure 4-5h

     4.5.7 Password

The system password can be changed. First, enter the old 
password, then enter the new password and confirm the 
new password. The default System password is “123”. Touch 
“ZERO” to adjust the value. Press “MODE” to shift to the next 
position. Touch “LOG” to cancel; Push “MENU” to confirm and 
exit. (Figure 4-5i)
 
 

 

   

 

Figure 4-5i

4.5.8 Key Setting
Set the default function of the “LOG” key from the home 
screen.  The function can be set to print or store the current 
displayed value. Press “ZERO” or “MODE” to select the proper 
setting.  Press “LOG” to cancel; touch “MENU” to confirm and 
exit. (Figure 4-5j)
 
 

 

 

 

   Figure 4-5j

4.5.9 RS232 Baurate
Adjust Baurate to available bits per second selection in Figure 
4-5k. 

                                       Figure 4-5k
4.5.10 Default Setting
If you make a selection that you feel has caused the gauge 
to operate improperly, you can restore it to the factory default 
settings. Carefully use this function! (Figure 4-5l)

 

   

 

 Figure 4-5l 

4.6 Language

Select between English, German and Chinese (Figure 4-6a)  

                                  

    

 

        Figure 4-6a

4.7 Calibration

Users can field-calibrate the gauge. First, enter the system pass-
word (Default is 123) by pressing the “ZERO” and  “MODE” keys. 
Then press “MENU” to confirm. (Figure 4-7a)
 

   Figure 4-7a                              

j

Calibration Point  

k

Current Value      

l

Standard Input Value

 

 

 

 

      Figure 4-7b

Load a standard force on the gauge. Wait a moment for the force 
to stabilize. The current value (2) should equal the standard force 
applied. 

If the values do not match, press “ZERO” and “MODE” keys to 
correct the standard input value (3).

Press “MENU” to enter the next calibration point. After any of the 
calibration points have been completed, touch “LOG” to exit the 
calibration mode. Then save the calibration or discard by pressing 
“Yes” or “No”.

After completing the calibration of the 5th point, the confirmation 
window will automatically ask to “Save and Exit” by selecting 
“Yes” or “No”. (Figure 4-7c)

  

 

 

Figure 4-7c

Press “ZERO” or “MODE” to select, then press “MENU”. If “Yes” 
is selected, “Calibration Complete!” is displayed.

 

6

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание FG-7000L

Страница 1: ...Operators should wear protection such as a mask and gloves in case pieces or components break away from the unit under test Whether the unit is ON or OFF DO NOT exceed the capacity of the sensor NEVE...

Страница 2: ...gher value is mea sured AutoPeak When Peak Time is up resets peak value automat ically First Peak Captures First Peak after drop ratio decrease has been detected Drop Ratio set in menus If no load cel...

Страница 3: ...to cancel or touch MENU to confirm and exit Figure 4 2c Figure 4 2c 4 2 2 Group When several test samples need to be measured the sam ples can be coded into groups The range is 01 99 When set to 00 b...

Страница 4: ...ENU to confirm and exit Figure 4 3b Single At the home screen pressing the LOG stores the cur rent displayed value If the default settings key is for storage See 4 5 8 key setting Series Continuous da...

Страница 5: ...rs may be set per Fig ure 4 5a 4 5b Figure 4 5a Figure 4 5b 4 5 1 Display Mode Two display modes may be selected Digital and Graphic Fig ure 4 5c Figure 4 5c 4 5 2 Display Direction Select the mode of...

Страница 6: ...can restore it to the factory default settings Carefully use this function Figure 4 5l Figure 4 5l 4 6 Language Select between English German and Chinese Figure 4 6a Figure 4 6a 4 7 Calibration Users...

Страница 7: ...1a The USB Port can be connected to a charger with the USB cable for charging the internal Ni MH battery and can be connected to a computer for uploading the measured values Connect the gauge and the...

Страница 8: ...sions W H B M 1kN 51 76 2 25 4 M12X1 75 2kN 5kN 10kN 20kN Capacity Dimensions mm D d L H SR DI M 10kN 32 8 14 16 16 24 5 3 M5 20kN 38 11 30 50kN 82 22 32 44 50 68 4 M8 100kN 200kN 100 28 36 48 120 80...

Отзывы: