background image

Method of operation 

By performing the following operations, it is possible to observe on the 

oscilloscope how the 10MHz output of the slave is phase-locked to the master 

and the rises of 1pps are aligned with microsecond accuracy.

 

1

 

Complete  the  above  connection.  Next,  10MHz  output  and  1pps  output  are  each 

monitored with an oscilloscope. 

1.1

 

Compare 10MHz with master and  slave, and when trigger is applied to master, 

slave signal flows to either right or left (frequency is different). 

1.2

 

When 1pps is compared between the master and the slave, the rising position is 

observed to be off by more than 1 microsecond. 

2

 

How to phase lock the slave module clock to the master module clock 

2.1

 

Press the “Lock” button on the Slave module 

2.1.1

 

Then  10MHz  will  phase  lock  within  a  few  seconds,  and  the  slave  signal  is 

observed to stop flowing. 

2.1.2

 

After the 10MHz phase lock is observed, the rise of 1pps is alighned within 5 

seconds with an accuracy of less than 1 microsecond. 

2.2

 

Press the “Lock” button on the Slave module again to “Unlock”. 

 

3

 

How to reset 

3.1

 

Slave module reset 

3.1.1

 

Press the “Reset” button on the Slave module. 