DC Power Supply
DC Power Supply
FPGA
PXIe-7821R
Measurement System
RTI-12308
SLSC-12202
Device Under Test
P3.7 as
Channel
Disconnect
Vsup
Vsup
DIO30
DIO31
DIO2
DIO1
DIO0
JR3 DIO30
JR3 DIO31
JR2 DIO2
JR2 DIO1
JR2 DIO0
JH1 Vsup_1
JH1 Vsup_0
High
Impedance
P3.6 as
Sourcing
Output
P0.2 as
Push-pull
Output
P0.1 as
Sourcing
Input
P0.0 as
Sinking
Input
Power
P3.5 as
Sinking
Output
JR3 DIO29
Vsup
Load
DIO29
Channel-Disconnect Switch
Channel-disconnect switch is used to connect or disconnect the digital input and output to the
front I/O connector.
Note
The channel-disconnect switch is controlled through software and the
switching rate is software-dependent.
Short-Circuit Protection
Digital output channels are protected against short-circuit faults.
Note
Short-circuit protection is based on hiccup-mode protection where the control
logic will disconnect the channel using the channel-disconnect switch once short-
circuit fault is detected. This channel will stay disconnected for one second before
the control logic attempts to enable this channel again automatically. However, if the
short-circuit fault is not removed, then this channel will continue to operate in
hiccup protection mode with invalid input or output signal.