NI SCXI-1379 Installation Instructions

Matrix expansion cable (to expand the columns of a matrix)

Matrix expansion plug (to expand the rows of a matrix)

Tip

NI recommends using 28 AWG wire (ribbon or discrete) for use with the SCXI-1379 

terminal block.

3. Connect Signals

To connect signals to the terminal block, refer to Figures 1 and 2 while 
completing the following steps:

1.

Prepare the signal wire by stripping the insulation no more than 7 mm.

2.

Remove the top cover screw.

3.

Unsnap and remove the top cover.

4.

Loosen the two strain-relief screws on the strain-relief assembly.

5.

Run the signal wires through the strain-relief opening. 

6.

Insert the stripped end of the wire fully into the terminal. Secure the 
wire by tightening the screw of the terminal. No bare wire should 
extend past the screw terminal. Exposed wire increases the risk of 
a short-circuit induced failure.

Note

When connecting signals to the SCXI-1379, refer to Table 1 to determine where to 

connect signals for the 8x32, 1-wire matrix topology.

7.

Tighten the two screws on the strain-relief assembly to secure the 
cables.

8.

Reinstall the top cover.

9.

Replace the top cover screw.

Содержание SCXI-1379

Страница 1: ... following conventions are used in this guide The symbol leads you through nested menu items and dialog box options to a final action The sequence File Page Setup Options directs you to pull down the File menu select the Page Setup item and select Options from the last dialog box This icon denotes a tip which alerts you to advisory information This icon denotes a note which alerts you to important...

Страница 2: ...damage when you handle the terminal block take the following precautions Caution Never touch the exposed pins of connectors Ground yourself using a grounding strap or by touching a grounded object Touch the antistatic package to a metal part of the chassis before you remove the terminal block from the package Remove the terminal block from the package and inspect the terminal block for loose compo...

Страница 3: ... the top cover screw 3 Unsnap and remove the top cover 4 Loosen the two strain relief screws on the strain relief assembly 5 Run the signal wires through the strain relief opening 6 Insert the stripped end of the wire fully into the terminal Secure the wire by tightening the screw of the terminal No bare wire should extend past the screw terminal Exposed wire increases the risk of a short circuit ...

Страница 4: ... 1 Terminal Block Top Cover 2 Signal Wires 3 Safety Ground Lug 4 Screw Terminals 5 Column Connector 6 Rear Connectors 7 Top Cover Screw 8 Thumbscrew 9 Column Connector 10 Strain Relief Screw 11 Strain Relief Bar 12 Strain Relief Opening 13 Row Connector 1 7 5 4 3 12 11 10 9 8 6 13 2 ...

Страница 5: ... EXPANSION CONNECTOR 176 POSITION CONNECTOR 176 POSITION CONNECTOR 128 POSITION ROW EXPANSION CONNECTOR COLUMN 0 C0 1 C1 2 C2 3 C3 4 C4 5 C5 6 C6 7 C7 8 C8 9 C9 10 C10 11 C11 12 C12 13 C13 14 C14 15 C15 32 C16 33 C17 34 C18 35 C19 36 C20 37 C21 38 C22 39 C23 40 C24 41 C25 42 C26 43 C27 44 C28 45 C29 46 C30 47 C31 1 Wire 0 R0 1 Wire ROW 1 R1 2 R2 3 R3 4 R4 5 R5 6 R6 7 R7 COLUMN EXPANSION CONNECTOR ...

Страница 6: ... 2 c8 COLUMN 8 c21 COLUMN 37 r3 ROW 3 c9 COLUMN 9 c22 COLUMN 38 r4 ROW 4 c10 COLUMN 10 c23 COLUMN 39 r5 ROW 5 c11 COLUMN 11 c24 COLUMN 40 r6 ROW 6 c12 COLUMN 12 c25 COLUMN 41 r7 ROW 7 c13 COLUMN 13 c26 COLUMN 42 c0 COLUMN 0 c14 COLUMN 14 c27 COLUMN 43 c1 COLUMN 1 c15 COLUMN 15 c28 COLUMN 44 c2 COLUMN 2 c16 COLUMN 32 c29 COLUMN 45 c3 COLUMN 3 c17 COLUMN 33 c30 COLUMN 46 c4 COLUMN 4 c18 COLUMN 34 c3...

Страница 7: ...d the rows of a 4x32 matrix using a matrix expansion plug The matrix expansion plug connects columns to expand the rows To expand the rows of a matrix complete the following steps 1 Connect two SCXI 1379 terminal blocks to the front connectors of the two adjacent SCXI 1130 modules 2 Tighten the top and bottom chassis screws on the back of the terminal block rear panels to hold them securely in pla...

Страница 8: ...terminals in one of the SCXI 1379 terminal blocks 2 Connect the other end of the matrix expansion cable to one of the ROW CONNECTION CABLE TERMINALS in another SCXI 1379 3 Connect the two SCXI 1379 terminal blocks to the front connectors of the two SCXI 1130 modules 4 Tighten the top and bottom chassis screws on the back of the terminal block rear panels to hold them securely in place Figure 4 Col...

Страница 9: ...e 5 and complete the following steps the SCXI 1130 should already be installed in an SCXI chassis 1 Plug the SCXI 1379 onto the front connectors of the SCXI 1130 2 Tighten the top and bottom thumbscrews on the back of the terminal block rear panel to hold it securely in place Figure 5 Installing the SCXI 1379 Terminal Block 1 Thumbscrew 2 Front Connectors 3 SCXI 1130 4 SCXI 1379 4 1 3 2 ...

Страница 10: ...ility Emissions EN 55011 Class A at 10 m FCC Part 15A above 1 GHz Immunity EN 61326 1997 A2 2001 Table 1 EMC EMI CE C Tick and FCC Part 15 Class A Compliant Note For EMC compliance you must operate this device with shielded cabling CE Compliance This product meets the essential requirements of applicable European Directives as amended for CE marking as follows Low Voltage Directive safety 73 23 EE...

Страница 11: ...nal Instruments trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products refer to the appropriate location Help Patents in your software the patents txt file on your CD or ni com patents 2003 2007 National Instruments Corporation All rights reserved 373691B Nov07 ...

Отзывы: