Chapter 2
Connecting Signals
©
National Instruments Corporation
2-3
Front Signal Connector
The NI PXI-4204 is a direct-connect device. The connection interface
consists of a 16-pin COMBICON connector and one SMB connector.
Figure 2-2 shows the pin assignments for the signals. Figure 2-3 shows the
front label of the NI PXI-4204, with each set of screw terminals labeled
according to the corresponding differential input signal. To connect a
signal to the NI PXI-4204, complete the following steps while referring to
Figures 2-2 and 2-3:
1.
Remove power from the signal lines. If this is not possible, complete
the following steps while referring to Figures 2-2, and 2-3.
a.
Remove the COMBICON connector from the NI PXI-4204.
b.
Attach the signal sources according to the instructions in steps 2
through 4.
c.
Ensure the NI PXI-4204 is powered on.
d.
Reinstall the COMBICON connector.
2.
Strip 7 mm (0.25 in.) of insulation from the ends of the signal wires.
3.
Insert the wires into the screw terminals.
4.
Tighten the screws to 0.5–0.6 N · m (4.4–5.3 lb · in.) of torque.
Connect a timing or triggering signal to the PFI0/CAL SMB connector
using a cable with an SMB signal connector.
Caution
The PFI0/CAL SMB connector is for low-voltage timing and calibration signals
only
. Voltages greater than ±15 V can damage the device.