Table 2. PXI/PXIe/PCI-5122 Front Panel Indicators
Label
Function
ACCESS Indicates the status of the PCI bus and the interface from the PXI/PXIe-5122 to
the controller.
ACTIVE Indicates the status of the onboard acquisition hardware of the PXI/PXIe-5122
Figure 5. PXI/PXIe/PCI-5122 9-Pin DIN Connector Pin Assignments
4
3
5
6
7
8
9
1
2
1. +5 V (Fused)
2. GND
3. Reserved
4. Reserved
5. Reserved
6. PFI 1
7. Reserved
8. Reserved
9. PFI 0
Configuring the PXI/PXIe/PCI-5122 in MAX
Use Measurement & Automation Explorer (MAX) to configure your NI hardware. MAX
informs other programs about which devices reside in the system and how they are configured.
MAX is automatically installed with NI-SCOPE.
1.
Launch MAX.
2.
In the configuration tree, expand
Devices and Interfaces
to see the list of installed
devices.
Installed devices appear under the name of their associated chassis.
3.
Expand your
Chassis
tree item.
MAX lists all devices installed in the chassis. Your default device names may vary.
Note
If you do not see your device listed, press <F5> to refresh the list of
installed devices. If the device is still not listed, power off the system, ensure
the device is correctly installed, and restart.
4.
Record the device identifier MAX assigns to the hardware. Use this identifier when
programming the PXI/PXIe/PCI-5122.
5.
Self-test the device by selecting the device in the configuration tree and clicking
Self-
Test
in the MAX toolbar.
NI PXI/PXIe/PCI-5122 Getting Started Guide
|
© National Instruments
|
9