Getting Started with the NI PCI-1426

2

ni.com

Measurement & Automation Explorer Help for NI-IMAQ

—Describes how to configure the 

NI-IMAQ driver software, NI image acquisition devices, and cameras using Measurement & 
Automation Explorer (MAX).

NI-IMAQ Help

—Contains fundamental programming concepts for the NI-IMAQ driver software 

and terminology for using NI image acquisition devices.

Safety Information

Caution

The following paragraphs contain important safety information you 

must

 follow when 

installing and operating the device.

Do 

not

 operate the device in a manner not specified in the documentation. Misuse of the device may 

result in a hazard and may compromise the safety protection built into the device. If the device is 
damaged, turn it off and do 

not

 use it until service-trained personnel can check its safety. If necessary, 

return the device to National Instruments for repair.

Keep away from live circuits. Do 

not

 remove equipment covers or shields unless you are trained to do 

so. If signal wires are connected to the device, hazardous voltages can exist even when the equipment is 
turned off. To avoid a shock hazard, do 

not

 perform procedures involving cover or shield removal unless 

you are qualified to do so. Disconnect all field power prior to removing covers or shields.

If the device is rated for use with hazardous voltages (>30 V

rms

, 42.4 V

pk

, or 60 V

dc

), it may require a 

safety earth-ground connection wire. Refer to the device specifications for maximum voltage ratings.

Because of the danger of introducing additional hazards, do 

not

 install unauthorized parts or modify

the device. Use the device only with the chassis, modules, accessories, and cables specified in the 
installation instructions. All covers and filler panels 

must

 be installed while operating the device.

Do 

not

 operate the device in an explosive atmosphere or where flammable gases or fumes may be 

present. Operate the device only at or below the pollution degree stated in the specifications. Pollution 
consists of any foreign matter—solid, liquid, or gas—that may reduce dielectric strength or surface 
resistivity. The following is a description of pollution degrees.

Pollution Degree 1—No pollution or only dry, nonconductive pollution occurs. The pollution has 
no effect.

Pollution Degree 2—Normally only nonconductive pollution occurs. Occasionally, nonconductive 
pollution becomes conductive because of condensation.

Pollution Degree 3—Conductive pollution or dry, nonconductive pollution occurs. Nonconductive 
pollution becomes conductive because of condensation.

Clean the device and accessories by brushing off light dust with a soft, nonmetallic brush. Remove other 
contaminants with a stiff, nonmetallic brush. The unit 

must

 be completely dry and free from 

contaminants before returning it to service.

You

 must

 insulate signal connections for the maximum voltage for which the device is rated. Do 

not

 

exceed the maximum ratings for the device. Remove power from signal lines before connection to or 
disconnection from the device. 

Caution

National Instruments measurement products may be classified as either Measurement 

Category I or II. Operate products at or below the Measurement Category level specified in the 
hardware specifications.

Содержание PCI-1426

Страница 1: ...are for developing applications NI Vision Builder for Automated Inspection NI Vision Development Module LabVIEW LabWindows CVI Microsoft Visual Basic Optional Equipment National Instruments offers a variety of products for use with the NI 1426 including the following RTSI bus cables for connecting the NI 1426 to other Vision Motion Control or data acquisition devices 15 pin cable assembly kits for...

Страница 2: ... introducing additional hazards do not install unauthorized parts or modify the device Use the device only with the chassis modules accessories and cables specified in the installation instructions All covers and filler panels must be installed while operating the device Do not operate the device in an explosive atmosphere or where flammable gases or fumes may be present Operate the device only at...

Страница 3: ...This category refers to measurements on hard wired equipment such as equipment in fixed installations distribution boards and circuit breakers Other examples are wiring including cables bus bars junction boxes switches socket outlets in the fixed installation and stationary motors with permanent connections to fixed installations Unpacking The NI 1426 ships in an antistatic package to prevent elec...

Страница 4: ...in Camera Link cable to the Base configuration Camera Link compatible camera Refer to your camera manufacturer documentation for specific instructions about how to connect the cable to your camera 10 Connect the Camera Link cable to the Camera Link connector on the NI 1426 front panel 11 Plug in and power on the computer The NI 1426 is now installed and the camera is connected Configuring the NI 1...

Страница 5: ...ch of the configuration tree 3 Expand the NI IMAQ Devices branch 4 Expand the NI PCI 1426 branch 5 Right click Channel 0 and select Camera 6 Select your camera from the menu If your camera is not in the menu verify that the appropriate camera file is installed in the NI IMAQ Data directory Note If you lose communication with the camera during an acquisition for any reason such as unplugging a cabl...

Страница 6: ...ial input threshold 0 2 V maximum RS 422 compatible Maximum quadrature count rate 5 MHz Note Camera Link cameras must transmit at a minimum of 20 MHz 1 Clocks Pixel clock frequency range 20 MHz to 50 MHz2 PCI Interface Theoretical max PCI bandwidth 133 MB s Memory Onboard memory 16 MB or 32 MB SDRAM depending on memory option purchased Serial Requirements Baud rates supported 300 600 1200 1800 200...

Страница 7: ... the following standards of EMC for electrical equipment for measurement control and laboratory use EN 61326 EMC requirements Minimum Immunity EN 55011 Emissions Group 1 Class A CE C Tick ICES and FCC Part 15 Emissions Class A Note For EMC compliance operate this device to product documentation CE Compliance This product meets the essential requirements of applicable European Directives as amended...

Страница 8: ... calibration you can obtain the calibration certificate for your product at ni com calibration National Instruments corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759 3504 National Instruments also has offices located around the world to help address your support needs For telephone support in the United States create your service request at ni com support and fol...