
©
National Instruments Corp.
15
NI 9201E/9221E
NI 9201E/9221E with Screw Terminal Isolation Voltages
Channel-to-channel........................... None
Channel-to-earth ground
Continuous ................................. 250 V
rms
,
Measurement Category II,
(Double insulation)
Withstand.................................... 2,300 V
rms
, verified by a 5 s
dielectric withstand test
Measurement Category II is for measurements performed on
circuits directly connected to the electrical distribution system.
This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example, 115 V for U.S.
or 230 V for Europe.
Caution
Do
not
connect the NI 9201E/9221E with screw
terminal to signals or use for measurements within
Measurement Categories III or IV.