Table 3. Autonomous Mode Capabilities
Capability
Description
Use the Open/Close FPGA
Reference VI functions to run
an FPGA VI and then
disconnect the USB cable.
Call the Close FPGA VI Reference Function without
aborting or resetting the FPGA VI before you disconnect
the USB cable.
1
After you disconnect the USB cable, any VI running on the
NI USB-7846R continues to run and collect data, which
can later be retrieved by reconnecting the USB cable and
re-opening the original FPGA reference.
Use Interactive Front Panel
Communication to run an
FPGA VI and then
disconnecting the USB cable.
The front panel indicators stop updating.
After you disconnect the USB cable, any VI running on the
NI USB-7846R continues to run. Restore the Interactive
Front Panel Communication by reconnecting the USB
cable.
Download the bitfile to flash
memory and set it to run when
loaded to FPGA.
Download the bitfile to flash memory and set it to run when
loaded to FPGA. The bitfile automatically starts running
whenever power is applied to the NI USB-7846R. To
collect data, you must reconnect the NI USB-7846R to a
host computer before powering off the device.
Refer to the
LabVIEW FPGA Module Help
for more
information about downloading an FPGA VI to the flash
memory.
Worldwide Support and Services
The National Instruments website is your complete resource for technical support. At
, you have access to everything from troubleshooting and application development
self-help resources to email and phone assistance from NI Application Engineers.
for NI Factory Installation Services, repairs, extended warranty, and
other services.
to register your National Instruments product. Product registration
facilitates technical support and ensures that you receive important information updates from
NI.
National Instruments corporate headquarters is located at 11500 North Mopac Expressway,
Austin, Texas, 78759-3504. National Instruments also has offices located around the world.
For telephone support in the United States, create your service request at
or
dial 1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the
1
The Host VI errors out if you do not call the Close FPGA VI Reference Function without aborting
or resetting the FPGA VI.