The manufacturing process operates under Class 1 of the Medical 

Devices Directive by fully complying with the essential 

requirements for the CE marking of this product. 

Flammability 

The Pro-Mat-Bario Mattress meets the requirements of BS7177:2008 for medium 
hazard.   

The interior materials meet the requirements of The Furniture & Furnishings (Fire 
Safety) Regulations 1988. 

User Instructions 

The Pro-Mat-Bario mattresses have been designed for use on mesh, slat & solid 
mattress supports and can be used on profiling & static bed bases, by patients in the ‘at 
risk’ user category

 with a weight guide of up to 294kgs / 46 stones (394kgs / 62 stones 

for Bario-Max) NOTE: You should check the Safe Working Load of the bed on which the 

mattress is to be used to ensure compatibility with mattress & user. 

Ensure that the surface of the bed is clear from debris before positioning the mattress – 

debris can damage the coating of the fabric. 

Frequent rotation of the foam mattresses is essential to ensure that loading is evenly 
spread across both ends in order to minimise foam core fatigue in use.  The mattress is 

for rotation only (head to foot & foot to head). It must not be turned over as pressure-
relieving qualities would then be compromised. 

The cover on this mattress is removable for cleaning purposes & to allow for inspection 
of the cover and the foam core for damage and contamination. It is recommended that 
frequent inspections of the cover are made to check for damage such as holes or cuts. 
Always ensure that, if removed, the cover is returned to the original foam core to avoid 
cross-contamination.  Using this cover on a different foam core may also render the 

warranty invalid. 

Instructions for use are screen printed onto the cover of the product for ease of 
reference. 

   

Indication of use based upon interface pressure readings – not a prescriptive criteria.  Always defer to clinician’s 

judgement.

 

 

Cleaning 

For general cleaning and superficial dirt, use a disposable wipe with a solution of warm 
water and neutral detergent.  Do not use abrasive or phenolic-based cleaners.  Do not 
mangle or iron. 

To disinfect in situ, use a 0.1% sodium hypochloride solution (1,000 ppm chlorine). 

A removable cover, identified as contaminated should be laundered at 65°C for not less 
than 10 minutes or at 95°C for not less than 3 minutes.  Do not mangle or iron.  Spin and 

tumble-dry at not more than 130°C.  Thoroughly rinse and dry the cover before re-use or 
storage.  Store in a cool dry area. 

Contaminated foam cores cannot be decontaminated and should be disposed of safely, as 

should any cover damaged by cuts, tears or ‘sharps’ penetration. 

Warranty

  

Covered against faulty manufacture and materials by replacement.   

Normal expected wear and tear in use is not covered.  

All warranties are subject to the terms and conditions of trading.   

Replacement value under warranty reduces proportionally to the period in use up to 3 years 

after purchase. 

If making a claim under warranty please quote the production identification batch number 
printed on the interior foam core. 

Storage & Care 

Store the mattress flat – do not fold 
Do not place heavy or sharp objects on the mattress surface when  in use / in store. 
Do not allow sharp objects to penetrate the cover coating (incl. hypodermic needles) 
Keep the mattress clean 
Regularly inspect the interior core and exterior cover of the mattress for damage or 
contamination. 
Do not store in damp conditions 

 