background image

 

 

 

SETUP 

Place  the  Bullet  Blender

®

  50  on  a 

stable,  level  lab  bench.  Carry  it  by 
grasping the bottom sides. Plug the AC 
power  supply  cord  connector  into  the 
AC  Input  Receptacle  on  the  right  side 
of  the  Bullet  Blender

®

  and  then  insert 

the plug into a wall outlet.  You should 
see  lights  on  the  time  setting  flash  for 
several seconds, indicating that the unit 
is receiving power. It is now set up. 

 

OPERATION 

To use your Bullet Blender 50, lift 

open  the  instrument  cover,  twist  the 
handle  of  the  sample  plate  cover 
counter  clock-wise  until  the  white 
arrow  is  in  the  unlocked  position  (see 
figure  below)  and  lift  the  sample  plate 
cover out of the instrument. 

 

 

The  figure  above  shows  labels  that 
indicate  if  the  sample  plate  cover  is  in 
the locked or unlocked position. Here, it 
is shown in the locked position. 
Insert  the  50  mL  sample  tubes  in  the 
protective sleeves and place the sleeves 

into the sample tube plate. 

The threads 

on the sample tubes must be dry and 
the  caps  must  be  screwed  on  very 
tightly. 

Use a minimum of 2 sleeves, so 

that  all  the  energy  from  the  Bullet 
Blender is not concentrated on a single 
tube. For best results, the sleeves should 
be  evenly  spaced.  Replace  the  cover 
and  rotate  it  to  lock  it  in  place.  Fold 
down the handle on the cover, and then 
close the lid. 

Set  the  time  and  the  speed  to  the 

button. At each three minute interval an 
LED  will  light  up,  indicating  its 
progress.  The  air  cooling  will  continue 
for  30  seconds  after  the  motor  times 
out. 

will stop the instrument. 

 

Example: 

 

 

The  figure  above  shows  the  BB50-AU 
model  Bullet  Blender  set  to  run  for  12 
minutes at a speed of 8 after running for 
6  minutes.  To  operate,  press  the 

LED  light  on the  number  1  will light 

DOMINIQUE DUTSCHER SAS

Содержание Bullet Blender 50

Страница 1: ...bsite www nextadvance com Click the SUPPORT button on the left hand menu bar and then on the appropriate link to the manual Blender will become an essential tool in your laboratory and we wish you suc...

Страница 2: ...ON THE BULLET BLENDER 50 Start Operation This product complies with European Low Voltage and EMC Directives Please dispose the test tubes and the BBX or BBY Bullet Blender in accordance with local reg...

Страница 3: ...e it is shown in the locked position Insert the 50 mL sample tubes in the protective sleeves and place the sleeves into the sample tube plate The threads on the sample tubes must be dry and the caps m...

Страница 4: ...of various beads as well as specific protocol information please refer to our website www nextadvance com For 50 mL tubes do not operate with more than 3 5g of sample or a total of 20mL of buffer tiss...

Страница 5: ...enly spaced when placed in the Bullet Blender If the 4 C light is not coming on frost from condensation may be blocking the air passage Remove the inner dry ice compartment and gently agitate it or re...

Страница 6: ...r is required Next Advance may also request documentation of proper maintenance if applicable Next Advance makes no other warranty express or implied with respect to its Products NEXT ADVANCE MAKES NO...

Страница 7: ...is not liable for any damages including but not limited to lost profits lost savings or other incidental or consequential damages arising from ownership or use of this product or for any delay in the...

Страница 8: ...50 8 D O M I N I Q U E D U T S C H E R S A S...

Отзывы: