4. Troubleshooting and Maintenance
General Data Buoy Maintenance
The following practices should be carried out for general CB-Series data buoy maintenance during
normal operation. As a general practice, measurement data should regularly be inspected, and
should be configured in WQData LIVE to detect any anomalies which may indicate that
maintenance is required.
1. Monitor the battery voltage and charge or replace as needed.
a. Battery voltage is reported as a standard metaparameter in WQData LIVE. Alarms can be
configured to monitor the charge.
b. If the charge of 12V sealed lead acid (SLA) batteries provided by NexSens (CB-A01/A05-X
battery packs) falls below ~10V, batteries should be charged or replaced.
CB-Series battery float charger kit
for recharging (30W for CB-A01-2 packs, 60W for
CB-A05-X packs).
ii. Battery replacement instructions are available
.
c. Batteries repeatedly falling below 10V should be replaced.
2. Carefully clean the buoy and connected devices as needed.
a. Clean off the solar panels with a soft towel using caution not to scratch the surface. Water is
normally sufficient for this, but a mild cleaning agent may be used if necessary. Do not use any
harsh or corrosive chemicals.
b. Ensure the vent on the data logger/data well lid is clear of all obstructions to avoid buildup of
high pressure gases in the data well from charging.
c. Inspect mooring lines and remove buildup from biofouling, especially if it is beginning to add
significant weight that is causing the buoy to sit low in the water. Replace any components
which are showing significant signs of wear.
d. Consult manufacturer’s instructions for cleaning of sensors and other connected devices.
NexSens CB-Series Battery
Float Charger Kit
47