NISE 3640M Series User Manual
Preface
Immunity Test
IEC 60601 Test Level
Compliance Level
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0 % UT (>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
0 % UT (>95 % dip in UT)
for 5 sec
0 % UT (>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
0 % UT (>95 % dip in UT)
for 5 sec
Power frequency
(50/60 Hz) magnetic field
3 A/m
3 A/m
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration
– Electromagnetic Immunity
The models in NISE 3640M series are intended for use in the electromagnetic
environment specified below. The customer or the user of the models
in NISE 3640M series should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2,5 GHz
3 V/m
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±2 kV ±4 kV ±6 kV contact
±2 kV ±4 kV ±8 kV kV air
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
±2 kV for power
supply lines
±1 kV for input/output
lines
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth