6300A Operator’s Manual
- 3 - NewTech Medical Limited
2 SYSTEM OPERATION PRINCIPLE
2.1
System-Block-diagram
Figure 2-1 6300A Diagram
Страница 1: ...L VENTILATOR 6300A NewTech Medical Limited 8400 Normandale Lake Boulevard Suite 920 Bloomington 55437 Minnesota U S A Tel 1 952 837 2688 Fax 1 952 400 8947 http www newtech medical com E mail info new...
Страница 2: ...LATION AND CHECK 8 3 1 Ventilator Component 8 3 2 Ventilator Installation 9 3 3 Check 9 4 OPERATION 11 4 1 Parameters setting 11 4 2 Operation 12 5 OPERATION METHODS OF VARIOUS VENTILATION MODES 13 5...
Страница 3: ...m the machine Always use a fresh bacterial filter and test circuit A hazard to the health of the service person may result z Hazardous voltages exist within this medical device Always observes appropr...
Страница 4: ...before putting the machine into operation and strictly follow all the procedures of operation Our company guarantees the quality of the product Any time and in any case when you need our help we are...
Страница 5: ...PONT 1 1 3 Ventilation Performance a Respiration frequency f 6 60 times min b I E Ratio 1 4 1 3 1 2 1 1 5 1 1 2 1 15 c Tidal volume 0 80 1200mL 15 d Ventilating volume 18L min 1 1 4 Monitor Performanc...
Страница 6: ...product should be stored in the room with temperature of 0 40 C relative humidity of no more than 80 good ventilation and without corrosion gas Note When the storage conditions are beyond the requirem...
Страница 7: ...6300A Operator s Manual 3 NewTech Medical Limited 2 SYSTEM OPERATION PRINCIPLE 2 1 System Block diagram Figure 2 1 6300A Diagram...
Страница 8: ...ue of the safety valve 5 1kPa 6kPa the safety valve can open automatically and the gas will be discharged from the safety valve The electromagnetic valve will be closed in expiration the pressure exer...
Страница 9: ...Fig 4 Table1 Figure 2 3 Table 2 1 Display symbol s implication Function TIDAL VOL Display the tidal volume value AIRWAY PRESSURE Airway pressure display Reflect the airway pressure change when patien...
Страница 10: ...airway pressure limit setting key LO PRESS Low airway pressure limit setting key POWER Power Indicator 2 2 3 Function key Figure 2 5 Table 2 3 Symbol Function A C Intermittent positive pressure ventil...
Страница 11: ...Airway High Pressure Alarm Display LO PRESS Airway Low Pressure Alarm Display Apnea Apnea Alarm Tidal Volume Tidal volume regulating knob Regulate ranging 0 80ml 1200ml 2min Eliminate alarm When alarm...
Страница 12: ...ilator Component Figure 3 1 1 Ventilator 2 Machine Arm 3 Power Cable For Ventilator 4 Check Valve 5 Oxygen Pipeline 6 Cart 7 Wheels 8 Bag 9 Water Trap 10 Lines For Measure Temperature 11 Pressure Samp...
Страница 13: ...power supply should be used The fuse of the 6300A ventilator is at the power socket position of the rear panel The fuse with the value of 2A can be replaced after the power cable is unplugged and the...
Страница 14: ...essure check Set the trigger pressure at 0 1cm H2O Pinch the simulated lung with both hands When the airway pressure is a little lower than the preset value trigger inspiration and the indicating ligh...
Страница 15: ...the key flashes and the revised part flashes At this moment press or key and set the respiratory frequency HI PRESS Airway upper pressure limit setting When pressing HI PRESS key the light at the lef...
Страница 16: ...gh When pressing the key the light at the left upper corner of the key flashes During A C every certain time provide double tidal volume once SIMV f 2 Synchronous intermittent mandatory ventilation Wh...
Страница 17: ...igher than the airway pressure peak value Figure 5 1 Control Ventilation Waveform 5 2 A Namely assist respiration It is featured by that the patient can control the respiration frequency while the tid...
Страница 18: ...ence breathe when Other parameter settings are same as that of A C b SIMV f 4 The command ventilation frequency is one fourth of A C ventilation frequency Others are same as that of f 2 Figure 5 3 SIM...
Страница 19: ...tient have spontaneous respiration select SPONT should be provided for the patient through facemask before offline so as to train the spontaneous respiration ability of the patient As a routine the me...
Страница 20: ...er to the above for the disinfectant method But the flow sensor should be paid special attention as it is precision and damageable part For disinfectant it is better to soak it in alcohol for an hour...
Страница 21: ...irection and it should not be installed in reverse Also the sensor of the inspiration circuit and that of the expiration circuit can t be mistaken Generally the expiration channel is easy to be pollut...
Страница 22: ...operation therapy or nursing of the patient Special the airway pipeline and power supply cable on the ground should not affect walking of person so as to avoid them being contacted and causing gas and...
Страница 23: ...off Switch on the power source Digital number tube doesn t flash Ventilator doesn t work Fuse melts Replace During operation ventilator stops suddenly indicating light goes out with sound alarm Power...
Страница 24: ...defects for a period of eighteen 18 months and accessories for a period of three 3 months The warranty period starts at the date of purchase The warranty is void if the ventilator machines have not b...