XPS-DRVP1
EDH0259En1010 – 08/10
20
11.0
Maintenance
11.1
Axis initialization after error
In case of malfunction, the XPS-DRVP1 driver board returns an error and the XPS
controller will automatically “kill” the axis. XPS website allows checking the type of
error(s).
Select FRONT PANEL/Driver Status to display the current XPS-DRVP1 driver board
status.
Following errors can occur on a XPS-DRVP1 driver board:
b: Broken fuse or Voltage out of range
c: Thermistor fault
d: Initialization error
e: (I
2
T) Dynamic error or Following error
g: Stage not connected
h: Inhibition input
i: Driver fault
Clicking on “Refresh” resets the XPS-DRVP1 driver board status.
After cancelling the error, the axis must be re-initialize.
This can be done through the XPS Website FRONT PANEL/Move page by clicking on
“Initialize”.