OPS-Q250, OPS-Q1000
Oriel
®
Quartz Tungsten
Halogen Lamp Power
Supplies
User's Manual
Family of Brands
–
ILX Lightw av e
•
New F
ocus™
Ophir
Corion
Richardson Grating
s™
Spectra-Phy sics
90074225 MOPS-Q, Rev B
Страница 1: ...s OPS Q250 OPS Q1000 Oriel Quartz Tungsten Halogen Lamp Power Supplies User s Manual Family of Brands ILX Lightwave New Focus Ophir Corion Richardson Gratings Spectra Physics 90074225MOPS Q Rev B...
Страница 2: ...NG THE LAMP ON OFF 18 4 3 LIGHT INTENSITY CONTROL 19 5 OPERATING MODES 20 5 1 POWERMODE 20 5 2 CURRENT MODE 20 5 3 INTENSITY MODE 20 5 3 1 Dosed Mode 20 6 POWER AND CURRENT OPERATION MODES 21 6 1 POWE...
Страница 3: ...el Connections 13 Figure 4 The LCD display of the OPS Q Model Power Supply 17 Figure 5 USB Connectorfor FirmwareUpdates 25 Figure 6 The LCD screen image displayedduring firmwareand LCD GUI interfaceup...
Страница 4: ...e as theOperatingMode of the OPS Q Power Supply 37 Figure 16 Idle FunctionSetup 39 Figure 17 Shutter Status Indicator on Display 40 Figure 18 Shutter Status Symbols 40 Figure 19 Manual Shutter Mode Sh...
Страница 5: ...mishandling could result in irreparable damage to the equipment and personal injury or death EUROPEAN UNION CE MARK The presence of the CE Mark on Newport Corporation equipment means that it has been...
Страница 6: ...ower sources before servicingthe equipment To avoid explosion do not operate this equipment in an explosiveatmosphere Qualifiedservice personnel should perform safety checks after any service If this...
Страница 7: ...wer supply A substitute line cord may not be rated for high current 1 5 LAMP HANDLING Read all informationand warnings provided with lamp Never touch any lamp or the reflector s inner surface with bar...
Страница 8: ...nel for easy navigation of power supply features LAMP and SHUTTER buttons located on front panel for manual control of lamp ignition and external shutter Icons on the LCD display show the ignition and...
Страница 9: ...90074225 MOPS Q QTH Lamp Pow er Supplies Page 9 2 2 DIMENSIONS Figure 1 Dimensionaldrawing of the OPS SeriesPower Supplies...
Страница 10: ...881 66882 66883 66884 67009 67011 60043 n a 6337 QTH 50W LONG FILAMENT 60000 Housing with 60090 Interface Kit 60043 66881 66882 66883 66884 67009 67011 60043 n a 6333 QTH 100W 60000 Housing with 60090...
Страница 11: ...e date When handling this power supply hold it from underneath the housing Never use any knobs buttons or cables to carry or locate the power supply WARNING Do not attempt tooperatethis equipment if t...
Страница 12: ...etting and monitoring various features of the power supply a Lamp OperatingMode b Manual TimedShutter Control DisplayFeatures c Saving loadingparticularsetups C SHUTTER This button is for manual contr...
Страница 13: ...o the main circuitry of the power supply There is no output from the power supply until the manual or external command to supply power to the lamp is received with the interlock conditionsatisfied 3 4...
Страница 14: ...this system Never open the OPS cover and attempt to work inside Use the power cord provided with the power supply Never alter the power cord in any way Be sure to periodically inspect the power cord a...
Страница 15: ...nectioncable provided by a third party vendor or made on the user s end could damage the power supply and lamp housingand pose an electrical shock risk to the user The connectorpin assignmentsare list...
Страница 16: ...L signal may be input to the TRIGGER INPUT connectoror a remote switch closure maybe wired to the connector G SHUTTER OUTPUT The BNC connector labeled SHUTTER OUTPUT outputs a TTL signal to an externa...
Страница 17: ...the power supply and related components lamp lamphousing cable s are preparedfor lampignition or displays Interlock Open indicating a problem with the system setup Similar error messages such as Cool...
Страница 18: ...the vertical menu Press the vertical menu button next to Pset Iset and use the knob to set this parameter when a red bullet appears next to the current value Then set Pmax Imax This is typically set...
Страница 19: ...d this value This value is to be used as the Flux Set set point iii If the recorded value shows value of 2 5uA 25uA or 250uA or 2500uA this indicates that the flux reading has overflowed with existing...
Страница 20: ...000 Q Series and 45 1000 W Oriel Research Lamp Housings adapters included and OPS model power supply While operating in Intensity Mode varying the Flux Set parameter allows the user to increase decrea...
Страница 21: ...djust the Pset and Pmax press the vertical menu button next to each parameter A red bullet in the box indicating that particular parameter can be changed will appear That parameter can now be changed...
Страница 22: ...er supply is being operated in Power Mode this indicates that the Imax setting in Current Mode is set too low for the lamp to reach the desired Pset If this error messageappears power down the lamp Th...
Страница 23: ...3 To adjust the Iset and Imax press the vertical menu button next to each parameter A red bullet in the box indicating that particular parameter can be changed will appear That parameter can now be ch...
Страница 24: ...s option can be changed by pressing the vertical menu key and using the external knob to select On or Off when a red bullet appears to the left of the current Audio setting LampType This is a read onl...
Страница 25: ...e or reversion if a previous version exists on the USB flash drive will occur The LCD GUI image file naming format will be imNN_XXX hex with NN being the image number 0 5 and XXX being the version num...
Страница 26: ...by the procedureexplainedpreviously in this section of the manual 8 INTERLOCK CABLES CONNECTIONS TO LAMP HOUSINGS A lamp interconnection cable must be purchased for use with a Research Lamp Housing b...
Страница 27: ...the previous page are used The use of a lamp housing interconnectioncableprovided by a third party vendor or made on the user s end could damage the power supply and lamp housing and pose an electrica...
Страница 28: ...esearch Lamp Housingfor use with an OPS Q Model Power Supply 8 3 SAFETY INTERLOCK Newport s Oriel Power Supplies have a safety interlock feature which must be satisfied before the power supply will po...
Страница 29: ...e to factors beyond the control of the power supply such as lamp aging ambient temperature fluctuations and electrode erosion For applications in which especially high light output stability is requir...
Страница 30: ...active area of the detector The predominant cause of dark current is thermionic emissions in which the detector outputs a current value as a result of ambient temperature To counteract this silicon ph...
Страница 31: ...ears the CoolTemp setting needs to be adjusted Sometimes external conditions such as an ambient operating temperature higher than typical room temperature 21 3 C can prevent the OPS Q Power Supply fro...
Страница 32: ...creases the frequency at which the photosensor must be returned for re calibration The OPS Q Power Supply is also specifically designed to control and maintain the operating temperature of the 71582 t...
Страница 33: ...couple the 71582 detector head to the Series Q Housing If an Oriel Products Research Lamp Housing is being used use the 68952 Adapter Kit to couple the 71582 detector head to the Research Lamp Housin...
Страница 34: ...n the rear of the 71582 Detector Head Then connect the 9 pin input connector on the other end of the 70062 to the 9 pin connectorheadon the rear panel of the OPS QPower Supply under thelabel markedTE...
Страница 35: ...S Q Power Supply Figure 16 The 70062 left and 70018 right requiredfor interfacingthe 71582 DetectorHead with OPS Q QTH Lamp PowerSupply 3 Ignite the lamp using either the Power or Current Operation Mo...
Страница 36: ...der the Setup indicator on the LCD display until System 2 appears with the correspondingvertical menu shown above Then press the vertical menu button to the right of the AutoSet indicator This automat...
Страница 37: ...supplied to the lamp in order to decrease the lamp s power output accordingly and increasing the Flux Set value will increase the current power supplied to the lamp in order to increase the lamp s pow...
Страница 38: ...stems where the 71582 TE Cooled Si detector is not coupled to the lamp housing and located after the shutter relative the light path In this type of setup the 71852 is not exposed to any light when th...
Страница 39: ...es are set by the user ExternalTTL control The shutter opens or closes based on an external TTL low or high signal ExternalTTL trigger The shutter begins a preset time or dose routine when triggered b...
Страница 40: ...will open or close based on the input of a TTL high or low signal from the Trigger Input connector on the back of the power supply If the shutter status symbol on the power supply display indicates th...
Страница 41: ...he button labeled SHUTTER on the front face of the power supply Figure 22 ManualShutter Button 11 3 TTL SHUTTERCONNECTION Any shutter that can be driven by a TTL signal may be controlled by the OPS po...
Страница 42: ...System 1 setup menu as shown below Press the vertical menu button next to the Shutter indicator When a red bullet appears next to the current Shutter setting use the knob to adjust to Pos or Neg as re...
Страница 43: ...lamp is being operated in Power Current and Intensity Modes 1 To operate the shutter in Timed Shutter Mode press the horizontal menu key under the Shutter indicator until Timed appears under the Shut...
Страница 44: ...next to the Back indicator to bring back the first menu shown on the previous page When the shutter is operated in Timed Exposure Mode new display options are available Pressing the horizontal menu k...
Страница 45: ...t the Flux Set Flux Multiplier Iidle and Imax parametersas necessary required 2 Press the horizontal menu key under the Shutter indicator until Dosed appears in the indicator with the correspondingver...
Страница 46: ...D screen shows that Dose and Total Dose are now available display options Total Dose is the accumlated Dose the lamp has been allowed to irradiate light onto the irradiationarea over multiple cycles 4...
Страница 47: ...gger Input connector while the cycle is in progress If the shutter is set to Manual mode and a TTL signal is sent to the Trigger Input connector the shutter remains open for the duration of the Low si...
Страница 48: ...nterlock Open Lamp housing connectioncable became loose or lamp housing door were opened when the lamp was ignited Ignition Failed Ignition failure during operation of the lamp A very low current draw...
Страница 49: ...e settings are automatically loaded Because of the versatility of the OPS power supplies Oriel recognizes that this unit may be utilized in a wide variety of projects Therefore the power supply was de...
Страница 50: ...ext to that Setup space The settings the power supply is currently operating under are now saved under the selected Setup space The following parameters are saved when a Setup is saved Lamp OperatingM...
Страница 51: ...and or damage to property Figure 28 LoadingSaved SettingsFrom Memory Loading a previously saved Setup is very similar to the methodused to save that Setup Press the horizontal menu button under the S...
Страница 52: ...gradually reducing the light output With tungsten halogen lamps the halogen gas effectively removes the deposited tungsten and returns it to the hot filament leaving the inside of the envelope clean a...
Страница 53: ...anel Use the scroll wheel on the power supply until Lamp Hours is selected Press the Display or CLEAR button in order to view the number of hours the lamp has been turned on Please note that Lamp Hour...
Страница 54: ...t ignited The Reset Lamp Hours feature was integrated into the power supply to allow a user to monitor lamp lifetime from installation or its first usage If necessary the monitoring of any time interv...
Страница 55: ...eter Resolution Current 0 01 A Safety Interlock Voltage 12 VDC GND OperatingMode Constant Current Power or Intensity OperatingTemperature 10 45 C Max Relative Humidity 80 non condensing Max Weight lb...
Страница 56: ...the cooling fan or vents of the lamp housing Fault light Display shows Ignition Failed Ticking sound heard from top of lamp housing during ignition cycle approximately once per second This messageis d...
Страница 57: ...supply If fault repeats contact Newport for RMA information Fault light Display shows Power SupplyFailed This messageis displayed when the power supplycannot be enabled for the lamp or there is a shor...
Страница 58: ...lux reading of the LIK LMP s 71582 detector head out of the recordablerange of the power supply 1 If the 71582 detector head is directlyconnectedto lamp housingswitch to the other Disk with Aperture i...
Страница 59: ...resthat the product ProductName Optical Power Supply Model Numbers OPS Q250 OPS Q1000 Type of equipment Electrical equipmentfor measurement control and laboratory use in industrial locations conformst...
Страница 60: ...used Pin 8 Not used Pin 9 Not used Figure 30 RS232 Connector Serial Port Setup 9600 Baud 8 bits 1 Stop No Parity No Flow Control Syntax ASCII Text Received commandlines are to be terminatedby CR Tran...
Страница 61: ...erlock isopen ESR none ESRXX HEX Bit5 Command Error Bit0 Operation Complete A responseof 64 indicatesthattheunit did notunderstandthe command Aresponse of 1 indicatesthatthe prior command has been pro...
Страница 62: ...mand for Hex valuedefinition Reset power supply to factory default RSTHRS none ESRXX SeeESRcommand for Hex valuedefinition Resetlamp hoursto 0 MODE i i 0 Power mode i 1 Currentmode ESRXX SeeESRcommand...
Страница 63: ...imit when lamp isoff P LIM iiii ESRXX SeeESRcommand for Hex valuedefinition Set power limit when lamp isoff STARTEXP none 0 for success Starts exposurer running STOPEXP none 0 for success Stops exposu...
Страница 64: ...Queries thetotal accumulatesopen exposuretime OPENTIME none hh mm ss th Queries theopen timefor exposures cycles OPENTIME hh mm ss th 0 for success Sets theopen timefor exposures cycles EXPOPEN none h...
Страница 65: ...ueries thefluxvalue FLUXMULT none ffff ff Queries thefluxmultiplier FLUXMULT ffff ff 0 for success Sets theflux multiplier FLUXCONTROL none 1 on 0 off Queries statusonflux control DETECTOR none ffff f...
Страница 66: ...to find range bit3 auto setup with cooler notready bit4 auto setup failed to seesignal bit5 auto setup detector saturated bit6 auto setup failed to find setpoint bit7 received command notunderstood b...
Страница 67: ...ices worldwide Every year the Newport Resource catalog is hailed as the premier sourcebook for those in need of advanced technology products and services It is available by mail request or through New...
Страница 68: ...RMA number must be obtained prior to shipment to Oriel Instruments This RMA number must appear on both the shipping containerand the packagedocuments Return the product to Oriel Instruments freight pr...
Страница 69: ...ed at Buyer s expense to Oriel facility provided that Buyer notifies Oriel in writing promptly after discovery of the defect or nonconformity and within the Warranty Period Products may only be return...
Страница 70: ...trade secrets of Newport and are for use only in the manner expressly permitted Newport claimsand reservesall rightsand benefitsafforded under law in the Programsprovided by Newport Corporation Newpor...