127
7
Appendix
7.1
Non Axis-Specific Error Messages
0, NO ERROR DETECTED
There are no error messages in the error buffer
3, OVER TEMPERATURE SHUTDOWN
The temperature inside the controller has exceeded 85
o
C. Under this condition, the controller
automatically stops any motion that is in progress and changes the status LED color to solid RED.
When the temperature goes below 75
o
C, the controller changes the status LED color to solid
GREEN. Motion related commands can once again be issued in this state.
6, COMMAND DOES NOT EXIST
The command sent to the controller is invalid. Please check the command syntax.
7, PARAMETER OUT OF RANGE
The parameter sent along with a command is out of range.
9, AXIS NUMBER OUT OF RANGE
The axis number specified is out of range.
10, EEPROM WRITE FAILED
Unable to save the controller’s settings in its non-volatile memory.
11, EEPROM READ FAILED
Unable to restore the controller’s settings from its non-volatile memory.
37, AXIS NUMBER MISSING
The command sent to the controller requires an axis number prefix. Please check the command
syntax.
38, COMMAND PARAMETER MISSING
The command sent to the controller requires a valid parameter. Please check the command syntax.
Содержание New Focus Picomotor 8743-CL
Страница 1: ...Picomotor Controller Driver User s Manual Model 8743 CL...
Страница 10: ...x Preface...
Страница 12: ...xii Preface Device Address Window 70 Internet Browser View 71 Command Syntax Diagram 73...
Страница 40: ...40 Multiple Units Stacking Interconnection Multiple Units Stacking Interconnection 8742 controllers shown...
Страница 59: ...59 Cycle Tab Cycle Motion...
Страница 60: ...60 Terminal Tab Command Line Terminal...
Страница 64: ...64 Options Window Connection Options Setup...
Страница 65: ...65 Properties Window USB and Ethernet Properties...
Страница 66: ...66 Controller Scan Scan RS 485 LAN...
Страница 67: ...67 Address Conflict Resolution Window Address Conflict Resolution...
Страница 70: ...70 Device Address Window...