Section 2 – Switch Control

2-1

Section 2
Switch Control

The four-pin electrical interface is used to control and
monitor switch position. The electrical connector is
Molex P/N 22-23-2041, and mates with Molex P/N 22-
01-3047 or equivalent.

Pin 4 is the position sensor output. The output
alternates between normal open contact and normal
close contact (low or high signal) depending on the
switch position. The position sensor is powered with
+5 VDC on pin 3. Pins 1, 2 and 3 are used for control
and power supply. Pin behavior depends upon the
electrical configuration of the user's switch. The
electrical configuration is described in the following
section, and summarized in Table 2.1, and Table 2.2.

1x2 Switch

Output 2

1x2 Switch

Output 1

Position Sensor

Position Sensor

Actuation Style

Pin 3

Pin 4

Pin 3

Pin 4

Non-Latching

2-Pin

+5 VDC

Low*

+5 VDC

High**

*

Normal open contact

*

Normal close contact

Table 2.1: Switch Control

Содержание MPSN Series

Страница 1: ...Fiber Optic Prism Switch USER S MANUAL...

Страница 2: ...trument returned transportation prepaid Repaired products are warranted for the balance of the original warranty period or at least 90 days Limitation of Warranty This warranty does not apply to defec...

Страница 3: ...ed for information only and product specifications are subject to change without notice Any change will be reflected in future printings 1999 Newport Corporation 1791 Deere Ave Irvine CA 92606 949 863...

Страница 4: ...itions 1 2 1 3 2 Terminology 1 2 1 4 Product Overview 1 4 Section 2 Switch Control 2 1 2 1 Non Latching Two Pin Control 2 2 Section 3 Specifications 3 1 3 1 Housing Dimensions 3 2 Section 4 Handling F...

Страница 5: ...2 Prism Switch 3 2 Figure 3 2 Housing Dimensions for 2x2 Prism Switch 3 3 Figure 4 1 Fiber Optic Component 4 2 List of Tables Table 2 1 Switch Control 2 1 Table 2 2 Switch Control 2 2 Table 3 1 Optica...

Страница 6: ...vi Preface...

Страница 7: ...oduct optical electrical environmental specifications and housing dimensions for the 1x2 and 2x2 Prism Switch Section 4 Handling Fiber Optic Components and Cables for the product are defined and also...

Страница 8: ...only used terms found in the manual 1 3 1 Safety and General Symbols Definitions The following is a definition of safety and general symbol used in this manual ________________________________________...

Страница 9: ...ed wavelength band Non Latching These style of switches require no power to maintain one position and a constant 5 VDC drive current to maintain the other position Polarization Dependent Loss PDL The...

Страница 10: ...itches are actuated electrically and require 5 VDC to switch position The switches are housed in a compact environmentally stable package designed for mounting on printed circuit boards or within a mo...

Страница 11: ...high signal depending on the switch position The position sensor is powered with 5 VDC on pin 3 Pins 1 2 and 3 are used for control and power supply Pin behavior depends upon the electrical configurat...

Страница 12: ...VDC Low GND 5 VDC 5 VDC High Normal open contact Normal close contact Table 2 2 Switch Control 2 1 Non Latching Two Pin Control Switches configured for non latching two pin control require no power t...

Страница 13: ...ut connectors Repeatability for 100 cycles at constant temperature Single mode only Measured at 1550 nm Table 3 1 Optical Specifications _______________________________________________________________...

Страница 14: ...ble 3 3 Environmental Specifications 3 1 Housing Dimensions The following illustrations Figure 3 1 and Figure 3 2 show the Housing Dimensions for 1x2 and 2x2 Prism Switch Figure 3 1 Housing Dimensions...

Страница 15: ...Prism Switch 10 1 25 6 7 2 2 54 TYP Pin 1 6 4 4 0 Aux Blind Mounting Hole M3X0 5 3 00 DP 2 Places 6 5 2 Places 4 0 6 0 16 0 Output 2 Output 1 3 3 7 8 11 4 20 0 40 0 61 0 26 0 3 3 Input 1 Input 2 32 0...

Страница 16: ...3 4 Section 3 Specifications...

Страница 17: ...ble strain relief boot Bending an optical cable near a strain relief boot is one of the easiest ways to permanently damage the optical fiber Avoid bending the optical cable over a sharp edge Avoid usi...

Страница 18: ...ables have been connected Figure 4 1 Fiber Optic Component 4 2 Storing Optical Connectors All switches are shipped with dust caps in place covering all optical connectors Optical connectors should rem...

Страница 19: ...ting process can permanently damage the connector Insert the appropriate connector ferrule into the adapter smoothly Do not allow the fiber tip to contact any surface If the tip accidentally contacts...

Страница 20: ...bles Optical power trending slowly increasing or decreasing is caused by the slow evaporation of alcohol trapped in the connection Continue to monitor optical power until it stabilizes If the loss is...

Страница 21: ...t com srvc service html Table 5 1 Technical Customer Support Contacts Contact Newport to obtain information about factory service Telephone contact number s are provided on the Service Form see next p...

Страница 22: ..._______________________ Reason for return of goods please list any specific problems ___________________ _____________________________________________________________________ _________________________...

Страница 23: ...been certified compliant with ISO 9001 by British Standards Institution P N 28989 01 Rev A IN 06999 07 00 Printed in the USA Newport Corporation Worldwide Headquarters 1791 Deere Avenue Irvine CA 9260...

Отзывы: