32
PART 8
APPROVALS INFORMATION
8.1 Electrical Compatibility (EMC)
This device comforms with requirements of EMC Directive 89/336/EEC, amended
by 93/68/EEC. This instrument complies with the following EMC Immunity
Standards as tested per EN 50082-2, 1995 (Industrial environment)
Phenomena
Test Specification
Basic Standard
Electrostatic
+/- 4 kV contact discharge
IEC 1000-4-2
Discharge
+/- 8 kV air discharge
Performance
Criteria B
Radio Frequency
27 - 1000 MHz
IEC 1000-4-3
electromagnetic
10 V/m
Performance
field.
80% AM (1 KHz)
Criteria A
Radio Frequency
900 MHz
IEC 1000-4-3
electromagnetic field.
10 V/m
Performance
Pulse modulated.
50% Duty cycle @ 200 Hz
Criteria A
Fast Transients
+/- 2 kV (ac mains)
IEC 1000-4-4
+/- 1 kV (dc, signal I/O)
Performance
5/50 ns Tr/Th, 5 KHz rep. freq.
Criteria B
Radio Frequency
0.15 - 80 MHz
IEC 1000-4-6
conducted
10 V/m
Performance
80% AM (1 KHz)
Criteria A
This instrument complies with the following EMC Emission Standards as tested per
EN 50081-1, 1992 (Residential, Commercial and Light Industrial)
Phenomena
Frequency
Limits
Basic
Range
Standard
Radiated
30-230 MHz
30 dB_V/m at 10 m
CISPR 22
Emission
230-1000 MHz
37 dB_V/m at 10 m
Class B
quasi peak
Conducted
0.15-0.5 MHz
66-56 dB_V quasi peak
CISPR 22
Emission
0.5-5 MHz
56 dB_V quasi peak
Class B
5-30 MHz
60 dB_V quasi peak
Safety
This device conforms with Low Voltage Directive 73/23/EEC, amended by
93/68/EEC. The following LVD requirements have been met to comply with
EN 61010-1, 1993 (Electrical equipment for measurement, control and
laboratory use)
1. Pollution Degree 2
2. Installation Category II
3. Double Insulation
4. Class II Equipment (90-240 Vac Powered Units)
Class III Equipment (12-36 Vdc Low Power Option)
8.2 FCC
This device complies with Part 15, Subpart B, Class B of the FCC rules