background image

 
To zero instrument: 
1.  Disconnect the sensor. 
2.  Make sure instrument is not in an electrically noisy environment and is 

undisturbed.  

3. Press “Zero” and “Start”. Wait until “Zeroing completed successfully” 

appears.  For more details, see main manual section 3.5.5. 

 

 

Thermal Sensors 

 

2.1. Use of 843-R with Thermal Type Sensors  

 

To set type of laser being used: 
1.  With the sensor attached, enter the Setup screen. 
2.  Press  the  Down  arrow  until  Laser  is  highlighted  and  press  the  Enter 

arrow. 

3.  Select the appropriate laser wavelength and press the OK key. 
4.  This will be saved automatically as the laser setting to work with. 
 
To choose manual or automatic ranging in power measurement: 
1.  With the sensor attached, enter the Setup screen. 
2.  Press  the  Down  arrow  until  Range  is  highlighted  and  press  the  Enter 

arrow. 

3.  Select the appropriate manual range or AUTO and press the OK key. 
4.  This will be saved automatically as the laser setting to work with. 
 
To choose power or energy measurement: 
In the main measurement screen press the Mode key. Select power or 
Energy and press the OK key. 
 

2.2. Other Settings and Saving the Startup Configuration 

 

1.  When measuring power, you can set the period over which to Average 

the measurements.  

2.  When  measuring  energy  you  can  set  the  threshold  level  in  order  to 

screen out false triggers.  

3.  The present settings are automatically saved for the next time 843-R is 

turned on. 

 

 

2.3 Power or Single Shot Energy Measurement 

 

Warning: 

 

Do not exceed maximum sensor limits for power, energy, power density and 
energy  density  as  listed  in  tables  5  and  6  in  section  9  of  the  main  manual. 
Otherwise, there is a risk of damaging the absorber. 
 
To use the 843-R to measure laser power: 
1.  Press the Mode key and select Power. Press the OK key. 
2.  In  the  Setup  screen  Press  the  Down  arrow  until  Range  is  highlighted 

and press the Enter arrow. Select AUTO or one of the manual ranges. 
The correct manual range is the lowest one that is larger than the  

expected  maximum  power  of  the  laser.  Note  that  the  range  will  be 

displayed in the lower right corner of the main display. 

 
To use the 843-R to measure laser energy: 
1.  Press the Mode key and select Energy. Press the OK key. 
2.  The  energy  mode  is  manual  ranging.  In  the  Setup  screen  Press  the 

Down arrow until Range is highlighted and press the Enter arrow. The 
correct  range  is  the  lowest  one  that  is  larger  than  the  pulse  energy 
measured.  Note  that  the  range  will  be  displayed  in  the  lower  right 
corner of the main display. 

3. When the 843-R screen flashes "READY", on and off, fire the laser. 
 
To select an analog needle type display for the 843-R: 
Press  the  “Display”  key.  Select  “Needle”  and  press  the  OK  key.  The 
display will now show a simulated analog needle type display. 
 
To expand the bargraph scale ±5x about the present reading: 
1.

 

From the main power measurement screen press the "Zoom” button.  

2.

 

Press the "Zoom” button again to return to the ordinary bargraph. 

 
To subtract background and set current reading to zero: 
1.

 

From the main power measurement screen press the "Offset” button. 

2.

 

Press "Offset" again to cancel.  

 
 

3

 Photodiode Sensors 

 

3.1. Selecting Chosen Wavelengths 

 
To set type of laser being used: 
1.  With the sensor attached, enter the Setup screen. 
2.  Press  the  Down  arrow  until  Laser  is  highlighted  and  press  the  Enter 

arrow. 

3.  Select the appropriate laser wavelength and press the OK key. 
4.  If  the  wavelength  you  want  is  not  among  the  wavelengths  in  the  six 

wavelengths listed press the Modify key. 

5.  Use  the  Right  and  Down  arrows  to  adjust  the  wavelength  as  desired. 

Then press the OK key. 

 
To choose manual or automatic ranging or dBm in power measurement: 
1.  With the sensor attached, enter the Setup screen. 
2.  Press  the  Down  arrow  until  Range  is  highlighted  and  press  the  Enter 

arrow. 

3.  Select the appropriate manual range or AUTO (or dBm for logarithmic 

scale) and press the OK key. 

4.  This will be saved automatically as the laser setting to work with. 
 
 

3.2. Other Settings and Saving the Startup Configuration 

 

1.  You can set the period over which to Average the measurements.  

2.  You can set the removable Filter setting to OUT for more accuracy and 

a  wider  wavelength  range;  you  can  set  it  to  IN  to  measure  higher 
powers.  

3.  The present settings are automatically saved for the next time 843-R is 

turned on. 

 
 

3.3. Power Measurement 

 

Warning: 

 

Do not  exceed  maximum sensor  limits  for  power,  energy,  power  density 
and energy density as listed in tables 5 and 6 in section 9 of the manual, 
Sensor  Specifications.    Otherwise,  there  is  a  risk  of  damaging  the 
absorber. 
 
To expand the bargraph scale ±5x about the present reading: 
1.

 

From  the  main  power  measurement  screen,  press  the  “Zoom” 

button.  

 

2.

 

Press the “Zoom” button again to return to the ordinary bargraph. 

 
To subtract background and set current reading to zero: 
1.  From the main power measurement screen press the "Offset” button.

 

2.  Press "Offset" again to cancel. See Section 5.4.5 of the main manual for 

details. 

 
 

 Pyroelectric or Photodiode Energy Sensors 

 

4.1 Zeroing Instrument against Sensor 

For  most  accurate  calibration,  you  should  zero  the  pyroelectric  sensor 
against the 843-R it is being used with. Proceed as follows: 
Make sure the sensor is in a quiet environment and not subject to pulsed 
radiation.  Plug  sensor  into  843-R  and  turn  on.  Enter  the  Setup  screen. 
Press  “Zero”  and  “Start”.  Wait  until  “Zeroing  completed  successfully” 
appears.  For more details, see main manual section 3.5.5. 
 

4.2. Changing Chosen Wavelengths 

1.  With the sensor attached, enter the Setup screen. 
2.  Press  the  Down  arrow  until  Laser  is  highlighted  and  press  the  Enter 

arrow. 

3.  Select the appropriate laser wavelength and press the OK key. 
4.  If  the  wavelength  you  want  is  not  among  the  wavelengths  in  the  six 

wavelengths listed press the Modify key. 

5.  Use  the  Right  and  Down  arrows  to  adjust  the  wavelength  as  desired. 

Then press the OK key. 

 

4.3. To Set Type of Laser Being Used 

To set type of laser being used: 
1.  With the sensor attached, enter the Setup screen. 
2.  Press  the  Down  arrow  until  Laser  is  highlighted  and  press  the  Enter 

arrow. 

3.  Select the appropriate laser wavelength and press the OK key. 
4.  This will be saved automatically as the laser setting to work with. 

Отзывы: