background image

Rev 1.00-1-

15-Mar-2013 yp 

Initial Revision  
Rev 1.24-1-

27-Aug-2014 yp 

Added 843-R-USB+note in 

 
 

4.4. Other Settings and Saving the Startup Configuration 

 

1.  You can set the period over which to Average the measurements.  
2.  You can set the Energy Range to measure in. Select the lowest range 

that is higher than the maximum expected energy measurement.  

3.  Set the Threshold to screen out noise that may be seen as false triggers 

and measured as energy pulses. 

4. If the sensor is a diffuser sensor, set diffuser to IN or OUT. 
5.  Select “Pulse Length” and choose the shortest pulse length setting that 

is longer than your laser’s pulse length.  

 

 Warning:  

Incorrect readings will result if pulse length is not set up correctly.  
 

4.5. Energy or Average Power Measurement 

 

Warning: 

 

Do not exceed maximum sensor limits for power, energy, power density 
and energy density as listed in tables 5 and 6 in section 9 of the manual. 
Otherwise there is a risk of damaging the absorber. 
 
With the pyroelectric sensor, you have been supplied a test slide with the 
same coating as on your pyroelectric detector.  You can also obtain this 
slide from your dealer.  You should use this slide to test the damage 
threshold    with    your    laser    pulses.   If the slide shows damage, then 
either enlarge your beam or lower the laser energy until damage is no 
longer seen. 
 

4.6. To choose Energy or Average Power Measurement 

In the main measurement screen press the Mode key. Select power or 
Energy and press the OK key. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Newport Corporation 

1791 Deere Avenue, Irvine, CA, 92606, USA

 

Phone. (949) 253-1694; (800) 222-6440 x31694  

www.newport.com

 

Rev 1.34-1/17.02.16/yp Newport P/N 1J06059 

 

 

843-R / 843-R-USB 
Power Meters 
 

843

-R

 / 843

-R

-USB

 Quick R

eference

 

 

 

Quick Reference

 

Getting Started 

 

Note:

  Any  mention  of  the  843-R  throughout  this  quick 

reference refers to the 843-R-USB as well. 
 
The  843-R  is  equipped  with  "soft  keys."  That  is,  the 
functions  of  the  keys  change  as  indicated  by  the  legend 
above each key. 

 

To connect sensor to the 843-R meter: 
Insert the 15 pin D type connector of the measuring sensor 
cable  into  the  socket  marked  "Sensor  Input"  on  the  rear 
panel of the 843-R meter. 

 

To switch the 843-R on: 
Briefly press the on/off switch (bottom-most key). The unit 
will  switch  on,  and  the  display  will  appear.  Note  that  the 
sensor must be plugged in before the unit is switched on. 
The  backlight  for  the  843-R’s  LCD  can  be  configured  to 
toggle between full, half, and low illumination. This toggling 
will be performed by briefly pressing the on/off switch after 
the  843-R has been  switched on. To switch the  843-R  off, 
press the on/off switch and hold it for ~ 2 seconds until the 
display blanks.  
 
The 843-R automatically saves the current for next power 
up. 

 

Enter the Setup screen to update the following: 
1.   Select “Show Settings” and enter. Set to Yes to display 

sensor settings in the measurement screens. Set to No 
to hide the sensor settings and show a larger graph. 

2. Select “Line Frequency” and enter. Set to 50Hz or 60Hz, 

depending on the electrical power grid of the area that 
you are in. 

3.  Select Date and enter. Scroll through and select Month, 

Day,  Year,  Hour  and  Minutes  with  the  Right  arrow. 
Change the selected item with the Down arrow. 

 