34 Computer
Interfacing
COMMAND EXECUTION:
The controller interprets the commands in the order they are received and
execute them sequentially. If a set of commands have to be executed closer
to each other, these commands can be sent to the controller simultaneously
by creating a command string with semicolon (;) used as a command
separator. The command string length should not exceed 50 characters. In
the example shown below, a command string was created with semicolon
separating 5 queries. The controller responds to this command string with a
response that has 5 values using a comma (,) as a separator.
COMMAND STRING:
LAS:OUT?;LAS:SET:I?;LAS:I?;LAS:SET:MDI?;LAS:MDI?
INSTRUMENT RESPONSE:
0,5540,0,33.94,0.00
COMMAND TERMINATION:
All commands sent to the instrument must be terminated by <Carriage
Return><Line Feed>.characters. All responses sent out by the instrument are
terminated by the same characters.
NOTE
For compatibility, some commands have optional entries; such as
LASer:LDI
, which
can also be
LASer:I
. Options are shown in Table 3 as parenthesis; such as
LASer:LDI (I)
. After Table 3, each command and query is detailed.
Содержание 500B Series
Страница 2: ...Laser Diode Drivers User s Manual Model 500B Series...
Страница 7: ...vi Preface This page is intentionally left blank...
Страница 23: ...22 General Information This page is intentionally left blank...
Страница 27: ...26 Getting Started This page is intentionally left blank...
Страница 47: ...46 Computer Interfacing This page is intentionally left blank...
Страница 51: ...50 Software Application This page is intentionally left blank...
Страница 59: ...58 Principals Of Operation This page is intentionally left blank...