16
o
CAL
– Compass calibration
▪
CALc –
compass calibration
(CI model only)
▪
Corr –
accuracy correction of
active measurement parameters
*not shown in the menu if currently selected
1
– available for CI model only.
Accuracy correction
If the System produces any measurement errors that lie
beyond unit specification, it may be an indication that unit
requires calibration similar to one conducted at
production level. To do accuracy correction, please select
Corr
from
CAL
menu.
Perform correction according to required measurement
parameters by pressing Action button, see table below:
Active mode of
measurement
(
Std
)
Correction
parameter
Condition
S, Sd, d
Distance
Install unit at 1 meter
1
from flat target
2
C, E
Azimuth and
elevation
Put unit in a level plane
and aim it to the North
direction
3