power well within 4 to 5 turns on each adjuster.
5) Observe the power out of the fiber as you slow-
ly adjust coarse
θ
x
. Turn the adjustment screw,
passing through the power peak and then return to
the maximum power seen on the power meter.
Remove the hex driver.
6) Repeat the same procedure as step five,
for the coarse
θ
y
adjustment.
7) Return to coarse
θ
x
and readjust, then
readjust the coarse
θ
y
until a maximum output is
seen on the power meter.
8) Once you have maximized power with the
coarse
θ
x
and
θ
y
adjustments, use a 1/16" driver to
adjust the fine
θ
x
and
θ
y
screws, on the side of the
stage, to further maximize coupling. The power
exiting the fiber will exceed 0.5 mW when properly
optimized.
If you are having problems adjusting the align-
ment to reach an output power of 0.5 mW, please
call us for assistance.
6
Fiber-Pigtail Manual revd 1/20/00 12:32 PM Page 5
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com