6 • Operation
NEW FOCUS, Inc.
oscilloscope and you want a positive output, use the
scope’s inverting x(-1) function.
Figure 1:
Model 1580-A
photoreceiver
module
Handling Precautions
Whenever handling the photoreceiver, make sure to
follow these precautions:
•
Prior to handling the detector or making
connections, be sure to ground yourself
adequately—even small electrostatic discharges
could permanently damage the detector. A ground
strap provides the most effective grounding and
minimizes the likelihood of electrostatic damage.
•
Do not over torque the microwave K-connector.
Excessive torque can damage connectors.
•
Make sure the optical connector is clean and
undamaged before connecting it to the detector
module.
Connecting the Power Supply and Bias
1.
Prior to handling the detector, ground yourself
with a grounding strap to prevent electrostatic
damage to the receiver.
Optical Input
Off On
Batt
Chk
Bias
Monitor
FC/PC connector
for fiber optic input
Bias-monitor port—output
is equal to photodiode current
times 1000
Ω
, for 1 mV/
µ
A.
Power switch
2.00 (50.8)
2.25 (57.1)
2.00
(50.8)
Output K-connector
3.14 (79.8)
Power
connector
1580-A Rcvr revA.fm Page 6 Friday, May 10, 2002 3:07 PM