TTL I/O OPERATIONAL TRIGGER DEFAULT SETUP
TRG [ <trigger setup> ]
Set/Query TTL I/O Operational Trigger default input configuration.
TTL I/O Operational Trigger is set to <trigger setup>.
<trigger setup> =>
FT
= Foot switch trigger (falling edge)
FH
= Foot switch hold (falling edge start, rising edge start)
F2
= Foot switch reverse (rising edge start/stop)
LE
= Level trigger (rising edge start, falling edge stop)
ST
= Start only trigger (falling edge start)
T2
= Start only trigger reversed (rising edge start)
SP
= Stop only trigger (falling edge stop)
P2
= Stop only trigger reversed (rising edge stop)
TTL I/O DIRECTIONAL CONTROL INPUT SETUP
DIN [ 0 | 1 ]
Set/query directional control input setup
Settings => 0 = Falling edge: Infuse, Rising edge Withdraw
Use this setting with the CBL-TTL-1, reciprocating pump cable, to create a 2 pump
continuous infusion system
1 = Falling edge: Withdraw, Rising edge Infuse
Use this setting with the CBL-TTL-1, reciprocating pump cable, to create a 2 pump
dual pumping system
PUMP MOTOR OPERATING TTL OUTPUT CONFIGURATION
ROM [ <on-off> ]
Set/query Pump Motor Operating TTL output configuration (TTL pin 7)
Settings => 0 = Output is logic high only when the pump motor is operating (pumping).
1 = Output is logic high when the pump motor is operating (pumping) or when the
Pumping Program is executing a pause timer.
SET LOCKOUT MODE
LOC [P] [ <on-off> ]
Set/query lockout mode. Lockout mode sets Auto-Run mode when used with Sub-Programs. See sec.
13, “Sub-Program Selection Input” for more information.
8.5.4
General Control and Status Commands
TTL I/O OUTPUT SETTING
OUT <n> <TTL level>
E { 1 |2 | 3 | 4 | 5 } <TTL level>
Expansion Port Connector
Sets TTL level on user definable output pin on the ‘TTL I/O’ connector.
<n> Indicates pin number on ‘TTL I/O’ connector
Valid value: 5 (Program Output pin)
E Sets pin on Expansion Port I/O connector on PCB.
Query is not applicable.