16
New Brunswick Scientific Co., Inc.
User’s Guide
Freezer
Socket
Alarm Plug
(external view)
Alarm Plug
(internal view)
WARNING!
Hazardous voltages must not be connected to the remote alarm socket.
Maximum rating is 24 volts, 1 amp.
Figure 3: Alarm Monitoring Socket & Plug
E
L
N
E
L
N
3A
250V
3A
250V
1
2
3
1
2
3
Normal Operation
Alarm Condition
Pin 1 (E)
Closed Circuit
Pin 2 (L)
Closed Circuit
OPEN
Pin 3 (N)
OPEN
Closed Circuit
The provided plug can be connected to either a central monitoring system or a
remote alarm.