24
M1288-0070
U360 Ultra-Low Temperature Freezer
User’s Manual
The following is a list of all operational based alarms:
•
Door open alarm
•
Power failure alarm
•
Electrical fault
•
Sensor failure
o
Cabinet temperature sensor
o
Ambient temperature sensor
o
Air-Cooled Condensor temperature sensor
o
Cascade Condenser temperature sensor
Each alarm is identified in plain English and displayed on the main screen for the
duration of the alarm or fault condition. All alarms, including a time/date stamp of
their occurrence, are also stored in the alarm log, found in the
DATA
menu screen.
4.4.1 Modifying
Alarm
Functions
1.
To modify the alarm functions, press the
PROG
menu key, then the
ALARMS
menu key. The alarm programming screen will be displayed:
Figure 5: Alarm Programming Screen
HELP
SAVE
EXIT
ALARM SOCKET DELAY:
30 MIN
TEMPERATURE ALARM DELAY:
60 MIN
DOOR OPEN ALARM DELAY:
1 MIN
AUDIBLE ALARM MUTE:
30 MIN
ALARM PROGRAMMING
2.
Using the
S
and
T
directional keys, select the alarm function to modify.
3.
Use the
W
and
X
directional keys to select among the preset options for
each alarm feature. (
Details are provided in Sections 4.4.2 through 4.4.7
.)
4.
Press the
SAVE
menu key to save your changes.