3.1 Physical
Location
WARNING!
Do not attempt to lift the Excella E-24/24R by yourself. Always ask for
assistance or use a lifter or other suitable equipment when raising or
handling the unit.
CRUSH WARNING!
It is preferable to lift the Excella E-24/24R from the sides of the unit,
but if you lift it from the front and back, be sure to use the finger holes
in the back lip (toward the bottom of the unit). If these finger holes are
not used, the back lip can pinch fingers and hands.
The surface where you place the E-24/24R should be smooth, level and sturdy, and must
be able to accommodate 200 pounds.
3.2 Environment
The shaker is designed to operate optimally in the following ambient conditions:
•
10º to 35°C
•
20 to 80% Relative Humidity (non-condensing)
3.3 Electrical
Requirements
The E-24/24R can be equipped to run on:
100 Volts, 50/60 Hz, 1500 VA maximum
120 Volts, 60 Hz, 1500 VA maximum
230 Volts, 50 Hz, 1500 VA maximum
In all cases, voltage variations must not exceed
±
10%.