NEW BRUNSWICK SCIENTIFIC Excella E-24 Скачать руководство пользователя страница 1

 

 
 

                                   

                

Excella E-24/24R 

Benchtop  

Incubator Shakers

  

 

 

 

 

 

     

MANUAL NO:  M1352-0050 

Revision A 

March 27, 2006 

 
 
 
 
 

 

 

 
 
 
 
 
 
 

 

 NEW BRUNSWICK SCIENTIFIC CO., INC. 

BOX 4005 

 44 TALMADGE ROAD 

 EDISON, NJ  08818-4005 

Telephone:  1-732-287-1200  

 1-800-631-5417   

Fax:  732-287-4222 

 Telex:  4753012 NBSCO  

Internet: http://www.nbsc.com 

 E-mail:  [email protected]

 

 

 

 
 
 
 

Содержание Excella E-24

Страница 1: ...AL NO M1352 0050 Revision A March 27 2006 NEW BRUNSWICK SCIENTIFIC CO INC BOX 4005 44 TALMADGE ROAD EDISON NJ 08818 4005 Telephone 1 732 287 1200 1 800 631 5417 Fax 732 287 4222 Telex 4753012 NBSCO In...

Страница 2: ...2 Excella E 24 24R Incubator Shakers M1352 0050 User s Guide...

Страница 3: ...Road Shanghai 200235 P R China Tel 86 21 6484 5955 or 5966 Fax 86 21 6484 5933 E mail nbschc online sh cn THE NETHERLANDS New Brunswick Scientific BV Kerkenbos 1101 6546 BC Nijmegen P O Box 6826 6503...

Страница 4: ...mage and personal injury can occur Please read the entire User s Guide before attempting to use this unit Do not use this equipment in a hazardous atmosphere or with hazardous materials for which the...

Страница 5: ...nt reflect our commitment to continuing product development and improvement Manual Conventions NOTE Notes contain essential information that deserves special attention CAUTION CAUTION messages appear...

Страница 6: ...Canada and for 1 year elsewhere Our obligation under this warranty is limited to repairing or replacing the instrument or part thereof which shall upon our examination prove to be defective The warra...

Страница 7: ...LLATION OF PLATFORM 15 4 2 FLASK CLAMP INSTALLATION 16 4 3 ELECTRICAL CONNECTIONS 17 5 FEATURES 18 5 1 CONTROL PANEL 18 5 1 1 Keypad 18 5 1 2 User Interface Keys 19 5 1 3 LED Display 19 5 1 4 Status I...

Страница 8: ...8 6 REPLACEMENT PARTS 35 9 ACCESSORIES 36 9 1 1 Platforms 36 9 1 2 Flask Clamps for Universal Platforms 36 9 1 3 Replacement Clamp Hardware Kits 37 9 1 4 Test Tube Racks Other Accessories 37 10 DRAWI...

Страница 9: ...perated in the following ways Continuously at a set speed and temperature until user intervention In a timed mode run at a set speed time and temperature for a period of up to 99 9 hours after which t...

Страница 10: ...ers M1352 0050 User s Guide Test tube racks microplate holders and test tube rack holders are also available a universal platform is needed for all test tube racks and holders For further information...

Страница 11: ...24 24R by yourself Always ask for assistance or use a lifter or other suitable equipment when raising or handling the unit CRUSH WARNING It is preferable to lift the Excella E 24 24R from the sides of...

Страница 12: ...ing shipping is missing pieces or fails to operate properly please contact your sales representative and also fill out Customer Satisfaction Form 6300 included with your equipment and return it by fax...

Страница 13: ...e sure to use the finger holes in the back lip toward the bottom of the unit If these finger holes are not used the back lip can pinch fingers and hands The surface where you place the E 24 24R should...

Страница 14: ...pth 30 inches 76 2 cm Height 24 1 16 inches 61 1 cm Height with lid open 40 inches 101 9 cm NOTE Be sure to allow at least three inches 7 6 cm around shaker for ventilation access to power cord rear p...

Страница 15: ...BPLATFORM of the BEARING HOUSING These screws must be removed and set aside for use before a platform can be installed Figure 2 Platform Screw Locations 1 Using the 7 32 inch hex wrench provided remov...

Страница 16: ...oles with holes on the platform Secure the clamp in place using the flat Phillips head screws provided S2116 3051 10 24 x 5 16 inch Use Figure 3b to help you identify the proper screws as three differ...

Страница 17: ...the others aside NOTE One liter and larger flask clamps are fastened with 5 screws 4 3 Electrical Connections CAUTION Before making electrical connections be sure to check the following 1 If you have...

Страница 18: ...switch is conveniently located on the right side of the unit as you face the controls Figure 4 Control Panel 5 1 1 Keypad The keypad see close up in Figure 4a on the following page is the operator s c...

Страница 19: ...W DOWN These keys are used to adjust the setpoint of a displayed parameter up or down They also allow the user to enter the SET MODE for setpoint changes 5 1 3 LED Display The digital display on the c...

Страница 20: ...TIME Indicates that the timer is in operation The shaker can be programmed to run for a preset time from 0 1 to 99 9 hours The timer can be disengaged without stopping an ongoing run 5 1 5 Function I...

Страница 21: ...T STOP key is pressed again When a timed run ends the HRS indicator will blink Press the SELECT key and change to another function to turn off this indicator MUTE This feature is controlled by the SEL...

Страница 22: ...niversal Voltage Input Module on the rear panel then connect the other end to a suitable grounded receptacle 6 2 Starting the Shaker To initially start the shaker close the lid and turn the ON OFF SWI...

Страница 23: ...6 4 Checking Any Setpoint 1 Press SELECT until the desired indicator is illuminated 2 Briefly press either the V or W KEY to enter the SET MODE and display the current setpoint CAUTION Holding the V o...

Страница 24: ...e START STOP KEY The TIME INDICATOR will cease to flash and the display will read OFF 6 6 Alarm Functions The Excella E 24 24R shakers have an audible alarm that is activated at predetermined times as...

Страница 25: ...temperature display Indicated Temperature match the temperature at a given point or match the average of a series of points within the chamber Actual Temperature proceed as follows 1 Let the unit equ...

Страница 26: ...of the LED display 6 10 Speed Calibration To calibrate the shaking speed 1 Set the shaker to a speed that can easily be measured If you are using a strobe minimum speed should be 250 RPM 2 Compare the...

Страница 27: ...t may be cleaned using a damp cloth or any standard household or laboratory cleaner to wipe down its outer surfaces Do not use abrasive or corrosive compounds to clean this instrument as they may dama...

Страница 28: ...samples with high concentrations of biomass or involving organic matter or occurring in areas warmer than room ambient temperature should be exposed to decontamination solution for at least one hour...

Страница 29: ...is indicated on display close lid firmly On Off switch is not working call for service If you recently replaced a fuse it may not have been seated properly remove and reinstall the fuse carefully Defe...

Страница 30: ...r on the outside of the shipping container otherwise NBS Receiving will refuse to accept the shipment In addition you must also certify that the instrument being returned has been thoroughly cleaned a...

Страница 31: ...the front panel in place setting them aside for reuse 3 Open the front panel allowing it to lay flat 4 The fuses are located on the PC board see Figures 6 7a 7b which you will find on the right side...

Страница 32: ...7a Fuse Replacement Excella E 24 Figure 7b Fuse Replacement Excella E 24R WARNING The following procedures are provided for your information only Do not attempt to perform these service interventions...

Страница 33: ...om the bearing housing mounting plate These nuts hold the flexure arms in place 7 Slide the upper bearing housing off the shaft and set it aside 8 Use the hex wrench to loosen the four hex nuts on the...

Страница 34: ...exure arms to the bearing housing mounting plate with the four nuts previously removed 17 Reattach the E ring to the shaft 18 Reinstall the platform on the bearing housing with the four Allen head scr...

Страница 35: ...you keep on hand one spare motor drive belt and two each replacement fuses NBS offers a spare parts kit that contains commonly needed replacement parts for your Excella E 24 24R see Table 2 below Tabl...

Страница 36: ...meyer flask Dedicated Platform 5 M1194 9908 2 8L Fernbach flask Dedicated Platform 4 M1194 9932 Utility carrier with cushioned crossbars M1194 9909 Utility tray with non skid rubber surface M1194 9910...

Страница 37: ...ck for tubes 8 11 mm diameter 48 tube capacity M1289 0001 5 60 tube capacity M1289 0200 4 44 tube capacity M1289 0020 5 Adjustable angle Test Tube Rack for tubes 12 15 mm diameter 34 tube capacity M12...

Страница 38: ...form Capacity Angled Test Tube Rack Holder for user supplied test tube racks that are 4 5 in 10 13 mm wide and up to 15 in 38 mm long TTR 210 2 Angled Test Tube Rack Spacer Bar for use with TTR 210 to...

Страница 39: ...ibbon Cable 1 2 3 4 5 J2 DISPLAY BOARD KEY PAD M1324 7002 YELLOW WHITE BLACK RED TB002 YELLOW RED WHITE BLACK TB002 TB001 COMPRESSOR UNIT SEE DRAWING 2 FOR DIFFERENT VOLT OPERATION 11 12 15 AMP CB001...

Страница 40: ...9 8010 M1324 8002 M1352 8003 M1352 8002 M1324 8010 M1352 8006 load PINK BRN RED WHITE M1324 7003 REFRIGERATION OPTION M1324 7004 SSR BOARD Figure 9b Control Schematics Quadrant B J1 CONTROL BOARD 3 4...

Страница 41: ...DRAWING 2 FOR DIFFERENT VOLT OPERATION M1352 8001 HEATER 120Vac OPERATION HEATER 100Vac OPERATION M1352 8007 REFRIGERATION OPTION J4 20 Cond Ribbon Cable 1 2 3 4 5 J2 DISPLAY BOARD KEY PAD M1324 7002...

Страница 42: ...1324 8016 BLACK BLACK M1352 8002 M1324 8015 TRANSFORMER POWER 100V 120V N 230V TB001 OVER PRESSURE COND FAN SWITCH COMPRESSOR UNIT Figure 10b 100VAC 60 Hz Schematic TB001 230V 120V 100V N COND FAN COM...

Страница 43: ...M 100Vac 50Hz OPERATION M1352 8002 POWER TRANSFORMER M1324 8015 ORANGE M1352 8007 BLUE OVER PRESSURE SWITCH M1324 8016 BLACK BLACK BLACK M1352 8007 VIOLET 2 1 Figure 10d 120VAC 60 Hz Schematic TB001 C...

Страница 44: ...le Clamp Installation 16 3b Clamp Fastener 17 4 Control Panel 18 4a Keypad 19 5 ON OFF Switch Location 22 6 Accessing Fuses 31 7a Fuse Replacement Excella E 24 32 7b Fuse Replacement Excella E 24R 32...

Страница 45: ...b 100VAC 60 Hz Schematic 42 10c 100VAC 50 Hz Schematic 43 10d 120VAC 60Hz Schematic 43 11 Refrigeration Schematic 44 10 3 List of Tables Table Description Page 1 Fuses 31 2 Spare Parts Kit M1352 6000...

Страница 46: ...ing SETPOINT RETENTION All setpoints and operating status are retained in non volatile memory AUTOMATIC RESTART Automatic restart after power is restored indicated by flashing display DRIVE UniCentric...

Страница 47: ...ntly reduce maximum speed for any shaker We may be able to improve this maximum speed by using an alternative motor pulley Contact your NBS representative for more information 11 1 Certifications The...

Страница 48: ...48 Excella E 24 24R Incubator Shakers M1352 0050 User s Guide...

Страница 49: ...e shipment In addition you must also certify that the instrument being returned has been thoroughly cleaned and decontaminated A form for this purpose is provided on the following page you can photoco...

Страница 50: ...____________________ This equipment check all that apply New Product Biohazards Never used Not used Used but decontaminated with _________________________ Hazardous Chemicals Radioactive Materials Not...

Страница 51: ...Form 6300 12 D Declaration of Conformity 48 Decontamination 27 Decontamination Certificate 30 49 50 Dimensions Chamber 47 Overall 46 Platform 47 Disclaimer Notice 5 Double Girdle Clamps 16 Drawings Li...

Страница 52: ...ed Material Authorization Number 30 49 RMA 30 49 50 RS 232 46 S Safety 46 Safety Switch 9 Setpoint Retention 46 Shaking Speed 46 Shipping Damage 12 Space Requirements 14 Spare Parts 35 Specifications...