37
Figure 7: Fuse Holder Detail
0.160A/250V/T 1.60A/250V/T 10A/250V/T
5. Gently twist the black cap of the fuse. The fuse pops out.
6. Remove the bad fuse and replace it with a new one.
7. Replace the black cap and twist until tight.
8. Reinstall the front panel, securing it with the three screws previously set aside. Use a
Phillips head (+) screwdriver to tighten the screws.
9. Close the door of the shaker, then reconnect the
POWER
CORD
to the electrical outlet.
8.4 Accessory Recorder Connector
To record speed and temperature, an accessory recorder can be installed on the back
panel of the shaker. The recorder should have the following capabilities:
•
Two channels, required for Speed and Temperature. Each channel should have
signal conditioning which accepts 0-5 volt input.
•
The pin-out diagram and scale below identifies the application.
A mating connector is required on the recorder cable (not supplied). This is a 9-pin male
D subminiature connector—AMP Amplimite HDP-20 series or equivalent.)
Figure 8: 9-Pin Recorder Connector
5 4 3 2 1
9 8 7 6
Pin Number
Signal Name
Scale
6
2
Speed
Ground
1V = 100 RPM
7
Temperature
1V = 20°C
3 Ground
New Brunswick Scientific Co., Inc.
User’s Guide