Rev. 01 on 20/01/2017
18/24
Ca p.11
- EL E CT ROM A GNE TI C COM PAT IBIL ITY
Table 201-Electromagnetic emissions declaration
Guidance and manufacturer's declaration - electromagnetic emissions
The MAGNETO BIOLIFE device is intended for use in the electromagnetic
environment specified below. The user must ensure it is used in this
environment.
Emissions test
Conformity
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 2
The MAGNETO BIOLIFE device must
emit electromagnetic energy to perform the
functions for which it was created.
Electronic equipment nearby could be
affected.
RF emissions
CISPR 11
Class B
The MAGNETO BIOLIFE device is
suitable for use in all establishments,
including domestic and directly connected
to the low voltage energy supply mains that
supplies buildings used for domestic
purposes.
harmonic
emissions IEC
61000-3-2
Not applicable
Voltage
fluctuations /
emission
flicker IEC
61000-3-3
Not applicable
Table 202-Declaration – Electromagnetic immunity
Guidance and manufacturer's declaration - electromagnetic immunity
The MAGNETO BIOLIFE device is intended for use in the electromagnetic
environment specified below. The user must ensure it is used in this environment.
Immunity test
Test level
IEC 60601
Conformity
level
Electromagnetic environment -
guidance
Electrostatic
discharge
(ESD)
IEC
61000-4-2
±6
kV
contact
±8 kV air
±6
kV
contact
±8 kV air
Floors should be in wood or
ceramic. If floors are covered
with synthetic material, relative
humidity should be at least 30%.
Table 204-Declaration – Electromagnetic immunity
Guidance and manufacturer's declaration - electromagnetic immunity
The MAGNETO BIOLIFE device is intended for use in the electromagnetic
environment specified below. The user must ensure it is used in this environment.
Immunity
test
Test
level
IEC
60601
Conformity
level
Electromagnetic environment - guidance
RF
3 Vrms 3Vrms
RF communications portable and mobile