16
wire or 12-wire 34 cm cables
provided with the system.
W
The use of cables or electrodes other
than the ones delivered with the
product might produce higher EMC
emissions and less EMC immunity.
W
The device cannot be used beside
or piled under other equipment.
If such usage is needed, check
the normal configuration.
W
The device can only be used in
healthy skin without wounds.
W
The device cannot be
used in a MRI room.
W
The device is not provided sterile
and should not be sterilized.
W
The device does not need installation,
maintenance or calibration.
W
The device and the
accessories should be regularly
checked by the user.
W
If the user wants to use the device
in combination to another device
connected to the patient, the user
should contact Neuroelectrics to
check the correct simultaneous use.
W
Starstim should not be used in an
MRI room or close to CT, diathermy,
RFID and electromagnetic security
systems such as metal detectors. In
the case that there exist RF emitters
(e.g. RFID), which might not be
visible, the device can potentially
be exposed to fields from these RF
emitters without the user’s awareness
and corrupts the signal acquisition.
If NIC detects that the signal is
very noisy, it will interference with
a higher Signal Quality Index.
W
The modification of the
device is not allowed.
W
If the device has not been used
during a long period of time, the
user should check visually that
there is no battery leakage.
W
The electrodes and wires or any
conductive part cannot touch any
other conductive part of any other
device including the ground.
W
The cap is intended to be on the
patient for less than 24 hours.
W
Keep out of reach from children and
anyone else who might swallow
electrodes, otherwise they may
cause injury to themselves.
W
Keep out of reach from children and
anyone else who might strangle
with the cables of the devices.
W
The result of the recordings must be
analysed by a doctor or specialist.
No self medication should be
done based on this results.
W
The result of the recordings is not
displayed in legal units or other units
within the meaning of Directive 80/181/
ECC. Therefore the device is not
considered to have a measuring function.
W
If the system encounters the
communication between Starstim
and the PC is failed, then the NIC
software or home interface will inform
the user accordingly. Additionally,
Starstim incorporates an internal buffer
that safeguards the information sent
from the device to the PC. If there’s
a problem of the communication
during a period of time, the device will
save the data in the internal buffer to
ensure than no EEG sample is lost.