NeuLog NUL-240 Скачать руководство пользователя страница 1

NEULOG AMMONIUM LOGGER SENSOR GUIDE

 

 

 

NeuLog ammonium logger sensor 
NUL-240 

 
The NeuLog ammonium ion sensor can be used for any science experiment which 
utilizes ammonium concentration readings such as in the fields of Environmental 
Science, Ecology, Chemistry, Biology, etc. 
 
Among hundreds of possible experimental subjects that can be studied with the NUL-
240 sensor are: chemical reactions, water quality, environmental health, small animal 
behavior, ecological studies, and many more. 
 
The ammonium sensor’s measurement units are: 

 

Milligrams per liter (mg/L): SI measurement unit for density of a solid 
dissolved in a liquid. 

 

Parts per million (ppm): The total amount of a specific molecule (ammonium 
for example) per million molecules in a sample.   
 

Ammonium sensor usage: 

 
The  ammonium  sensor’s  electrode  uses  a  PermaFil  (non-refillable),  ion-selective 
electrode.    The  reference  chamber  is  gel  filled  and  sealed;  therefore,  no  reference 
filling solution is required. 
 

Usage guide: 

 

Concentration range 

1M to 5x10

-6

0.01 to 18,000 ppm 

pH range 

4 to 10 pH 

Temperature range 

0 to 50°C 

Minimum sample size 

3mL in a 50 mL beaker 

Reproducibility 

±4% 

Electrode resistance 

1 to 4 megaohms 

 

Equipment required: 

 

 

Wash bottle with distilled or deionized water 

 

Several clean beakers 

 

1 mL, 10 mL and 50 mL pipettes (a 50 ml graduated cylinder can also be used). 

Technical notes:

 

 

 

All standard solutions should be prepared fresh. 

 

The standard solutions should be at the same temperature as the sample. 

 

Use ISA in all solutions. 

Note:

 ISA is added to all solutions to ensure the samples and standard solutions 

have the same ionic strength; the ratio is 1:50 

 

Electrode preparation: 

 
1. 

Remove  the  protective  cap  encasing  the  ammonium  probe’s  electrode.  

CAUTION:  Do  not  touch  the  PVC  membrane  at  the  end  of  the  probe  with 
your fingers. 

2. 

Rinse the electrode with deionized water and blot dry.  Be sure not to rub as this 
will harm the sensor. 

3. 

Soak  the  electrode  in  deionized  water  for  10  minutes,  then  in  a  diluted 
ammonium  standard  solution  (either  0.0001  M  or  1  ppm)  for  two  hours  until 
ready for use. 

4. 

For  preparing  the  1  ppm  standard  solution,  mix  one  drop  of  the  1000  ppm 
standard solution with 20 drops of ISA and add deionized water up to a total of 
40 ml solution  (assuming  one  drop  equals  0.04 ml). It  is  also  possible to  use  a 
micropipette if available. 

 

Offsetting the sensor: 

 
Be sure to offset the ammonium sensor once daily during periods of regular usage. 
 
1. 

Prepare the electrode as described in the “Electrode preparation” guide above. If 
you have  already prepared the electrode in the last three days and stored it as 
described  in  the  "Electrode  storage"  section  then  you  do  not  have  to  repeat  it 
before offsetting the sensor. 

2. 

Prepare  the  offsetting  solution:  Mix  1  ml  of  the  ISA  solution  with  0.5  ml  of  the 
1000 ppm standard solution. Add deionized water up to a total of 50 ml. You will 
get a 10 ppm solution.  

3. 

Rinse the electrode with deionized water, blot dry and place in the beaker. 

4. 

After  a  stable  reading  has  been  reached, 

press the  “Start/Stop”  button  located 

on the sensor’s faceplate for 4 seconds. 

5. 

Rinse the electrode once more with deionized water and blot dry. 

6. 

The ammonium sensor is now ready for use. 

 
 

Отзывы: