NEULOG PH LOGGER SENSOR GUIDE

 

 

 

Prior to initial use the Neulog pH Sensor needs to be calibrated, please review the 
“Sensor Calibration” section of this document starting on page 2 for detailed 
instructions.  Connect the pH Sensor to a USB-200 Module.  The USB-200 Module 
then connects to a computer via USB to mini USB cable.  Please note that you cannot 
plug the pH probe directly to the computer.   
 
Resident PC software or browser base software can be downloaded for free at 
www.neulog.com/dowload as well as a full software user guide.  Make sure not to 
download and install both types of software as they will conflict on the computer. 
 

Procedure: 

 

1. 

Remove pH Sensor probe from the plastic storage solution bottle. 

2. 

Gently rinse tip of pH Sensor’s probe thoroughly with distilled water. Be 
careful not to physically come in contact with the glass bulb. 

3. 

Calibrate the pH Sensor.  (Follow the ‘Sensor Calibration” guide in this 
document) 

4. 

Install the Neulog software 

5. 

Connect the USB-200 Module to the PC or Mac 

6. 

Connect the pH Sensor to the USB-200 Module (they directly plug together).   

7. 

Open the Neulog software. 

8. 

Once a pH Sensor logo appears on the left side of the screen the probe has 
been automatically identified and you can begin experimentation. 

9. 

If the pH Sensor is not automatically identified then click the search sensors 
icon to find the sensor. 

10. 

Select the “On-line experiment”, this will open a graph below. 

11. 

Click on the “Module setup” button located on the pH Sensor icon in the 
Module window to change the pH probe settings if need be. 

12.  Click on the experiment set up button to change the experiment settings if 

need be (Experiment duration for example). 

13.  The pH Sensor will give a live reading in the box to the left of the screen 

when running. 

14. 

To run an experiment and collect data click “Run experiment”. 

15. 

To end data collection early, click “Stop experiment”  

 
 
 

Tablet, smart phone device 

 

Materials needed: 
 

 

NUL-206 PH Sensor 

 

WIFI-201 WIFI module 

 

BAT-200 Battery 

 
Prior to initial use the Neulog pH Sensor needs to be calibrated, please review the 
“Sensor Calibration” section of this document for detailed instructions.  Your pH 
Sensor needs to be connected to a WIFI-201 Module.  The WIFI-201 Module will 
create a closed Neulog wifi network which will stream the Neulog data to a device of 
your choosing.  Once your device is wirelessly connected to the Neulog network you 
can run experiments and collect data through a browser of your choosing. 
 

Procedure: 

 

1. 

Remove pH Sensor probe from the plastic storage solution bottle. 

2. 

Rinse tip of pH Sensor’s probe thoroughly with distilled water.  Be careful 
not to physically come in contact with the glass bulb. 

3. 

Calibrate the pH Sensor.  (Follow the ‘Sensor Calibration” guide in this 
document) 

4. 

Connect the pH Sensor directly to the left side WIFI-201 (no wires required). 

5. 

Connect BAT-200 to the right side of the WIFI-201 

6. 

Although not required, we recommend plugging the BAT-200 to an outlet 
using a USB to micro USB charger (such as a typical cell phone charger).  
The wifi201 will run for 

X

 minutes (depending on the sensor) without being 

plugged in. 

7. 

For further WIFI-201 instructions or the WIFI-201 quick start guide please 
visit www.Neulog.com/download. 

8. 

The WIFI-201 indicator lights will flash; take no action until the LED to the 
far left turns blue. 

9. 

Take your tablet or smart phone and go to Wi-Fi settings and select the 
Neulog network which matches the WIFI Module ID found on the back of the 
WIFI-201 device. (NeuLogXXXX for example) 

10.  Give your device 1-2 minutes to connect to the WIFI-201. 