2.5 SET
OPERATION
Press set from main menu
000.0 KPA
unt
zro
peak
set
The following display will appear:
Setup Menu
Leak Alrm
2.5.1 LEAK DETECTION
The UniMano 1000 is capable of calculating the change in pressure over a
selected period of time.
Press [F1] Leak to select Leak Detection and the following menu appears:
Leak in Seconds
15
30
45
60
Use the [F] keys to determine the time span for the period of time you wish to
measurement to occur. Time intervals are in 15 second durations up to a
minute.
Once selected the display will revert back to the main display.
At the end of the selected measuring period the UniMano will beep and flash
the change in pressure value (5) five times and then the value freezes on the
display.
1.00 PSI
15
Seconds
leak
In the above example the UniMano detected a 1 PSI leak over 15 second
period, on negative sign in front would indicate a increase in pressure of 1 PSI
over the same 15 Second interval, see below.
-1.00 PSI
15
Seconds
leak
UniMano 1000 Series
12