background image

Delta 1600 - Quick-Reference Manual | 

13

• 

Connect the manufacturer specific interface cable between the 
Delta 1600 (red shrouded banana jacks) and the AED under test.   

• 

From  the AED  Main  Menu,  press  ‘F1’(Arrhythmia  Menu)  and 
select  the  Arrhythmia  waveform.  (The  default  waveform  is 
“VFBC”)

• 

Select the desired Arrhythmia waveform by pushing F1, F2, F3 
or F4. 

• 

The selected Arrhythmia waveform will be displayed (flashing) 

as illustrated below.

• 

Turn the AED power ON and follow the manufacturer’s operating 
instructions and the audio response. 

Shock Advisory Test

WARNING

Before getting started, follow the 
manufacturers safety instructions and test 
procedures.
Defibrillator discharge voltages are 
potentially dangerous. To avoid potential 
shock hazards, observe all safety 
procedures for the safe handling of the 
instrument before attempting to test the 
defibrillator.

Содержание Delta 1600

Страница 1: ...External Defibrillator Tester Intended Use Delta 1600 AED Tester is a precision instrument that simulates various cardiac waveforms and measures discharged energy It assures AEDs detect the waveforms...

Страница 2: ...tended User 4 Environment 4 Introduction 5 Delta 1600 Features 5 Specifications 6 Energy Measurement 6 General Specifications 7 Part Numbers and Ordering Information 8 Preparation for Use 9 Initial Se...

Страница 3: ...ustomer Service The Service Return Form may be obtained at our website http www netechcorp us RMARequests or contact our Netech Customer Service Department Customerservice netechCorp us Contact Inform...

Страница 4: ...ation where it might come in contact with a patient The Delta 1600 is intended to be used within the published specifications Any unauthorized alteration or modifications will result in a hazardous co...

Страница 5: ...The Delta 1600 has a built in load resistance of 50 ohms to simulate the human body impedance To establish electrical connection between the AED and the Delta 1600 manufacturer specific interface cabl...

Страница 6: ...J Accuracy 1 of reading 1 Load Resistance 50 ohms 1 non inductive 1 H Sampling Time 20 uS Measurement Time Window 60 ms Pulse Width 60 ms Patient Simulator Waveform Simulation ECG NSR ECG Rate 60 70...

Страница 7: ...r Tachycardia VTAC Atrial Fibrillation AFIB General Specifications Temperature Range Operating 15 C to 35 C Storage 0 C to 50 C Display Type LCD with LED backlit Graphics display Power 7 4 Volt Li Ion...

Страница 8: ...1 ZOLL ZOLL 630 651 SURVIVAL Cardiac Science Survivalink Burdick Cardiovive 630 651 PHYSIO Physio Control Life Pak Medtronic 630 612 Philips HeartStart FRx 630 615 Philips and Laerdal HeartStart FR3 6...

Страница 9: ...Delta 1600 AED Tester Li Ion Rechargeable Battery Charger 110 220 V input User manual USB Jump Drive Before getting started it is important to get familiar with the control and operation of the unit...

Страница 10: ...lowing are the description of the Key Functions F1 F2 F3 F4 and These are interactive keys associated with respective function selections ESC Reverts to the previous menu AUTO Pressing AUTO key change...

Страница 11: ...y Charger Li Ion Battery Charger provided with the unit is specifically designed for the Delta 1600 When the battery is fully charged the red LED on the CHARGER changes to green Battery Delta 1600 inc...

Страница 12: ...Press F2 ECG for the ECG Rate and Amplitude Menu Initial Setup Introduction The Delta 1600 is simple to operate and requires minimum setup However the user must read the manual thoroughly for safe and...

Страница 13: ...ng F1 F2 F3 or F4 The selected Arrhythmia waveform will be displayed flashing as illustrated below Turn the AED power ON and follow the manufacturer s operating instructions and the audio response Sho...

Страница 14: ...AED and wait until the energy is discharged Once discharged the energy will be displayed as illustrated below If the response is no shock advised select a different arrhythmia waveform from the arrhy...

Страница 15: ...multiple discharges and helps preserve the battery life of the AED Select AUTO key and wait for the AED s response The default arrhythmia waveform is set to VFBC If shock advised discharge the AED The...

Страница 16: ...function can be selected from the following ECG selection menu Rate Selection The default ECG rate is set 90 BPM ECG rate can be changed in 10 BPM increments from 60 BPM up to 200 BPM and 20 BPM incr...

Страница 17: ...ival at our facility These warranties are made upon the expressed condition that 1 The purchaser promptly notifies Netech in writing of any non conformity with the above warranty including a detailed...

Страница 18: ...ling charges will be billed to the purchaser THIS WARRANTY EXCLUDES ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED ORAL OR WRITTEN INCLUDING WITHOUT LIMITATION WARRANTIES OF MERCHANTABILITY AND OR...

Страница 19: ...Delta 1600 Quick Reference Manual 19 Appendix A Netech warrants this instrument to the original purchaser as Current Revision R1 06 12 17...

Отзывы: