Delta 1600 - Quick-Reference Manual |
13
•
Connect the manufacturer specific interface cable between the
Delta 1600 (red shrouded banana jacks) and the AED under test.
•
From the AED Main Menu, press ‘F1’(Arrhythmia Menu) and
select the Arrhythmia waveform. (The default waveform is
“VFBC”)
•
Select the desired Arrhythmia waveform by pushing F1, F2, F3
or F4.
•
The selected Arrhythmia waveform will be displayed (flashing)
as illustrated below.
•
Turn the AED power ON and follow the manufacturer’s operating
instructions and the audio response.
Shock Advisory Test
WARNING
Before getting started, follow the
manufacturers safety instructions and test
procedures.
Defibrillator discharge voltages are
potentially dangerous. To avoid potential
shock hazards, observe all safety
procedures for the safe handling of the
instrument before attempting to test the
defibrillator.